Jove
Visualize
Contact Us
JoVE
x logofacebook logolinkedin logoyoutube logo
ABOUT JoVE
OverviewLeadershipBlogJoVE Help Center
AUTHORS
Publishing ProcessEditorial BoardScope & PoliciesPeer ReviewFAQSubmit
LIBRARIANS
TestimonialsSubscriptionsAccessResourcesLibrary Advisory BoardFAQ
RESEARCH
JoVE JournalMethods CollectionsJoVE Encyclopedia of ExperimentsArchive
EDUCATION
JoVE CoreJoVE BusinessJoVE Science EducationJoVE Lab ManualFaculty Resource CenterFaculty Site
Terms & Conditions of Use
Privacy Policy
Policies

Related Concept Videos

Radioactive Decay and Radiometric Dating02:48

Radioactive Decay and Radiometric Dating

40.0K
Radioactivity is a spontaneous disintegration of an unstable nuclide and is a random process, as all the nuclei in the sample do not decay simultaneously. The number of disintegrations per unit time is called the activity (A), which is directly proportional to the number of nuclei in the sample. The decay constant (λ) is an average probability of decay per nucleus in unit time.
40.0K
Instrument Calibration01:12

Instrument Calibration

1.1K
Instrument calibration is essential for ensuring that instruments produce accurate and consistent results. It is vital in manufacturing, healthcare, testing laboratories, and scientific research. Calibration processes are specific to each instrument and help enhance data accuracy. Each instrument has a unique calibration process tailored to its design and function to improve data accuracy.
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
1.1K
Calibration Curves: Linear Least Squares01:20

Calibration Curves: Linear Least Squares

5.1K
A calibration curve is a plot of the instrument's response against a series of known concentrations of a substance. This curve is used to set the instrument response levels, using the substance and its concentrations as standards. Alternatively, or additionally, an equation is fitted to the calibration curve plot and subsequently used to calculate the unknown concentrations of other samples reliably.
For data that follow a straight line, the standard method for fitting is the linear...
5.1K
Calibration Curves: Correlation Coefficient01:10

Calibration Curves: Correlation Coefficient

5.5K
In a linear calibration curve, there is a value called the calibration coefficient, denoted by 'r,' which measures the strength and the direction of association between two variables. The correlation coefficient value ranges from −1 to +1. A value of +1 indicates a perfect positive linear correlation, −1 denotes a perfect negative correlation, and 0 implies no correlation between the two variables. A positive correlation value establishes that as one variable increases, the...
5.5K
Constant Volume Calorimetry02:41

Constant Volume Calorimetry

31.5K
Calorimeters are useful to determine the heat released or absorbed by a chemical reaction. Coffee cup calorimeters are designed to operate at constant (atmospheric) pressure and are convenient to measure heat flow (or enthalpy change) accompanying processes that occur in solution at constant pressure. A different type of calorimeter that operates at constant volume, colloquially known as a bomb calorimeter, is used to measure the energy produced by reactions that yield large amounts of heat and...
31.5K
Flame Photometry: Lab01:16

Flame Photometry: Lab

1.2K
In a flame photometer, when a solution like potassium chloride is aspirated into the flame, the solvent evaporates, leaving behind dehydrated salt. This salt dissociates into free gaseous atoms in their ground state. Some of these atoms absorb energy from the flame, leading to their excitation. The excited atoms return to the ground state, emitting photons at characteristic wavelengths. Because only electronic transitions are involved, the resulting emission lines are very narrow. The intensity...
1.2K

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Linearity Characterization and Uncertainty Quantification of Spectroradiometers via Maximum Likelihood and the Non-parametric Bootstrap.

Metrologia·2024
Same author

Incandescent lamp-based facility for measuring the long-term radiometric stability of spectrographs.

Applied optics·2022
Same author

Characterization and absolute calibration of an AERONET-OC radiometer.

Applied optics·2021
Same author

Improvements in the design of thermal-infrared radiation thermometers and sensors.

Optics express·2019
Same author

Invited Article: Advances in tunable laser-based radiometric calibration applications at the National Institute of Standards and Technology, USA.

The Review of scientific instruments·2018
Same author

The 1990 NIST Scales of Thermal Radiometry.

Journal of research of the National Institute of Standards and Technology·2017

Related Experiment Video

Updated: Mar 22, 2026

Laser-heating and Radiance Spectrometry for the Study of Nuclear Materials in Conditions Simulating a Nuclear Power Plant Accident
09:18

Laser-heating and Radiance Spectrometry for the Study of Nuclear Materials in Conditions Simulating a Nuclear Power Plant Accident

Published on: December 14, 2017

11.1K

"Once is Enough" in Radiometric Calibrations.

Gerald T Fraser1, Charles E Gibson1, Howard W Yoon1

  • 1Optical Technology Division, National Institute of Standards and Technology, Gaithersburg, MD 20899-8440.

Journal of Research of the National Institute of Standards and Technology
|April 26, 2016
PubMed
Summary

New quality systems for optical radiation measurement services streamline calibration. By focusing on process control and understanding, repetitive measurements are eliminated, improving quality and reducing costs.

Keywords:
calibrationefficiencyprocess controlquality measurements

More Related Videos

Split Point Analysis and Uncertainty Quantification of Thermal-Optical Organic/Elemental Carbon Measurements
10:22

Split Point Analysis and Uncertainty Quantification of Thermal-Optical Organic/Elemental Carbon Measurements

Published on: September 7, 2019

8.9K
Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
06:20

Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition

Published on: March 11, 2021

7.8K

Related Experiment Videos

Last Updated: Mar 22, 2026

Laser-heating and Radiance Spectrometry for the Study of Nuclear Materials in Conditions Simulating a Nuclear Power Plant Accident
09:18

Laser-heating and Radiance Spectrometry for the Study of Nuclear Materials in Conditions Simulating a Nuclear Power Plant Accident

Published on: December 14, 2017

11.1K
Split Point Analysis and Uncertainty Quantification of Thermal-Optical Organic/Elemental Carbon Measurements
10:22

Split Point Analysis and Uncertainty Quantification of Thermal-Optical Organic/Elemental Carbon Measurements

Published on: September 7, 2019

8.9K
Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
06:20

Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition

Published on: March 11, 2021

7.8K

Area of Science:

  • Optical physics and metrology
  • Measurement science and instrumentation

Background:

  • Development of a quality system within the Optical Technology Division.
  • Need to reassess existing calibration procedures for optical radiation measurement services.

Purpose of the Study:

  • To improve the quality and reduce the costs of optical radiation measurement services.
  • To revise calibration procedures by eliminating uninformative repetitive measurements.

Main Methods:

  • Implementing a quality system for optical radiation measurement services.
  • Instituting new procedures focused on controlling and understanding the measurement process.
  • Revising the first calibration program within the service to align with new procedures.

Main Results:

  • Successful development of an Optical Technology Division quality system.
  • Identification of opportunities to eliminate uninformative repetitive measurements.
  • Revision of calibration procedures in the first program.

Conclusions:

  • The revised calibration procedures enhance measurement quality and reduce costs.
  • Focusing on measurement process control and understanding is key to efficient calibration.
  • The implemented quality system provides a framework for continuous improvement in optical radiation measurement services.