Holmium Oxide Glass Wavelength Standards
1National Institute of Standards and Technology, Gaithersburg, MD 20899-8442.
Summary
Holmium oxide glass standards exhibit remarkable stability and minimal spectral variation over decades. Due to this long-term reliability, the National Institute of Standards and Technology (NIST) will discontinue recertification, simplifying traceability for users.
Area of Science:
- Spectroscopy and Metrology
- Materials Science
Background:
- Holmium oxide glass has been a trusted wavelength standard for over 40 years.
- These standards demonstrate minimal spectral variation across different batches and manufacturers.
- The National Institute of Standards and Technology (NIST) has a long history of certifying these samples.
Purpose of the Study:
- To evaluate the long-term stability and reliability of holmium oxide glass as a wavelength standard.
- To determine if continued recertification by NIST is necessary given the material's performance.
- To inform users about updated traceability procedures for holmium oxide glass standards.
Main Methods:
- Analysis of historical spectral data from NIST-certified holmium oxide glass samples.
- Review of NIST's extensive experience and data on holmium oxide glass stability.
- Assessment of the material's robustness against environmental factors like temperature and humidity.
Main Results:
- Insignificant spectral variation observed in holmium oxide glass standards over decades.
- No recorded instances of spectral shifts in certified bands.
- Holmium oxide glass is robust and insensitive to normal temperature and humidity fluctuations.
Conclusions:
- NIST will no longer recommend the recertification of holmium oxide glass wavelength standards.
- Traceability can be established through suppliers or end-users, reducing the need for NIST involvement.
- The long-term stability and reliability of holmium oxide glass justify this change in NIST's recommendation.
Related Concept Videos
UV–Vis Spectrometers
4.4K
The absorbance of UV and visible (UV–visible) radiations is measured using a UV–visible spectrophotometer. Deuterium lamps, which emit UV radiation, and tungsten lamps, which produce radiation in the visible region, are used as light sources in UV–visible spectrophotometers. A monochromator or prism is used for diffraction grating, i.e., to split the incoming radiation into different wavelengths. A system of slits is used to focus the desired wavelength on the sample cell.