Related Experiment Video
Updated: Mar 22, 2026

Transcorporal Artificial Urinary Sphincter Cuff Placement in a Case Requiring Revision for Urethral Atrophy
Published on: June 16, 2022
Maximum Urethral Closure Pressure Increases After Successful Adjustable Continence Therapy (ProACT) for Stress
Sarah H M Reuvers1, Jan Groen1, Jeroen R Scheepe1
1Department of Urology, Erasmus MC, Rotterdam, The Netherlands.
Objective:
To evaluate changes of the urethral pressure profile (UPP) after implantation of adjustable continence therapy (ProACT), a minimally invasive procedure in which 2 volume-adjustable balloons are placed periurethrally for treatment of male stress urinary incontinence. The working mechanism of the ProACT to achieve continence has not been fully understood. We hypothesized that successful treatment with ProACT improves urinary continence by inducing a significant increase in static urethral pressure.
Materials And Methods:
We included patients who underwent UPP before and after ProACT implantation. UPPs were initially performed with the Brown-Wickham water perfusion method and later with the T-DOC Air-Charged catheter method. Pre- and postoperative UPPs and International Prostate Symptom Scores were evaluated. UPP measurements of successfully (no or 1 precautionary pad per day) and unsuccessfully treated patients were compared.
Results:
Twenty-seven patients were included in the study; 23 patients were successfully and 4 patients were unsuccessfully treated. Maximum urethral closure pressure (MUCP) increased significantly from median 58.0 to 79.0 cmH2O in the successfully treated group (P = .001). Within the subgroup of unsuccessfully treated patients, MUCP did not change significantly (P = .715). The change in MUCP was statistically significantly different between the successful and unsuccessful group (P = .034). Total score of the International Prostate Symptom Scores did not change significantly after ProACT implantation (P = .097).
Conclusion:
Successful treatment with ProACT is associated with a significant increase of MUCP. This implies that increased static urethral pressure contributes to the working mechanism of the ProACT device to achieve continence.
Related Concept Videos
Urodynamic Studies: Uroflowmetry
Urinary Tract Calculi VI: Surgical Management
Urologic Endoscopic Procedure: Cystoscopic Examination
Urinary Tract Calculi III: Medical Management