A Modified QuEChERS-HPLC Method for Detection of Polycyclic Aromatic Hydrocarbons in Zebrafish Embryos Exposed to Fine Particulate Matter
Published on: June 13, 2025
High performance solid-phase extraction cleanup method coupled with gas chromatography-triple quadrupole mass
Fang Li1, Jing Jin2, Dongqin Tan1
1Key Laboratory of Separation Science for Analytical Chemistry, Dalian Institute of Chemical Physics, Chinese Academy of Sciences, No. 457 Zhongshan Road, Dalian 116023, China; University of Chinese Academy of Sciences, Beijing 100049, China.
Abstract:
A solid-phase extraction (SPE) cleanup method was developed to purify the sample extracts for the analysis of polychlorinated naphthalenes (PCNs) and dioxin-like polychlorinated biphenyls (dl-PCBs). Monodisperse magnesium oxide (MgO) microspheres and basic alumina were used as SPE adsorbents. Important parameters of the SPE procedure were optimized, including the amount of basic alumina and the type and volume of the washing and elution solvents. The optimized SPE cleanup method exhibited excellent purification performance for the removal of organochlorinated compounds, lipid compounds, sulfur, and pigments. Additionally, it was found that the retention activities of congeners differed with the number and position of the chlorine substituents in PCNs. In this study, an analytical method based on a combination of accelerated solvent extraction (ASE) coupled with SPE cleanup and gas chromatography-triple quadrupole mass spectrometry (GC-MS/MS) is proposed for the analysis of PCNs and dl-PCBs in complex samples (sediment, pine needle, and scallop samples). The analytical method demonstrates good linearity, acceptable recovery (63-148%) and precision (relative standard deviations less than 26%). The limits of detection (LODs) of PCN and dl-PCB congeners were in the range of 0.6-19.1pgg(-1) and 0.4-8.6pgg(-1), respectively. The PCNs and dl-PCBs levels in these samples ranged from 0.16 to 3.07ngg(-1) dry weight (dw) and from undetectable to 0.07ngg(-1) dw, respectively.
More Related Videos
09:04Identifying Per- and Polyfluorinated Chemical Species with a Combined Targeted and Non-Targeted-Screening High-Resolution Mass Spectrometry Workflow
Published on: April 18, 2019
04:17Analysis of Organochlorine Pesticides in a Soil Sample by a Modified QuEChERS Approach Using Ammonium Formate
Published on: January 20, 2023
Related Concept Videos
High-Performance Liquid Chromatography: Introduction
In HPLC, two phases play a critical role in the separation process:
High-Performance Liquid Chromatography: Elution Process
High-Performance Liquid Chromatography: Types of Detectors
High-Performance Liquid Chromatography: Instrumentation
Mass Spectrometry: Complex Analysis
GC–MS is a powerful hyphenated method commonly used in forensics and environmental...
Gas Chromatography: Types of Columns and Stationary Phases
For an analyte to remain on the column for a sufficient amount of time, it must exhibit some level of compatibility (or...