A Simplified (Screening) Approach to Check the Calibration Status of the Goldmann Applanation Tonometer
Nikhil S Choudhari1, Suresh Ramavath, Gunturu Rekha
1*VST Glaucoma Centre, Dr Kallam Anji Reddy Campus, L.V. Prasad Eye Institute, Hyderabad, Telangana †GMR Varalakshmi Campus, L.V. Prasad Eye Institute, Visakhapatnam, Andhra Pradesh ‡L.V. Prasad Eye Institute, Bhubaneswar, Odisha, India.
Background:
To investigate whether a lower-end calibration error check alone can identify a Goldmann applanation tonometer (GAT) with acceptable calibration error as per Asia-Pacific Glaucoma Society (APGS) and World Glaucoma Association (WGA) guidelines.
Methods:
This was a multicentric prospective study in a tertiary eye-care institutional setting. Seventy-six slit-lamp-mounted Haag-Streit GATs (Model AT 900 C/M) were included in the study. All GATs were checked for calibration error at 0, 20, and 60 mm Hg testing levels as recommended by the manufacturer. The ability of 2 screening criteria (calibration error within ±1 or ±2 mm Hg, respectively, at the 0 mm Hg testing level without using the calibration error check weight bar) to identify a tonometer within acceptable calibration limits defined by APGS (calibration error within ±2, ±3, and ±4 mm Hg at 0, 20, and 60 mm Hg testing levels, respectively) and WGA (calibration error within ±1 mm Hg at all testing levels) was determined.
Main Outcome Measure:
The negative likelihood ratio of the screening criteria RESULTS:: Five (6.6%) and 42 (55%) GATs were outside of the APGS and the WGA calibration limits, respectively. Negative likelihood ratios of the first and the second screening criteria at the APGS definition of calibration error tolerance were 0.11 and 0.25, respectively, and at that of the WGA definition of calibration error tolerance were 0.38 and 0.90, respectively.
Conclusions:
The screening criteria of calibration error within ±1 mm Hg at the 0 mm Hg testing level seem to be clinically useful to detect GAT with an acceptable calibration error as per the APGS guideline.
More Related Videos
Related Concept Videos
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Assessment of blood pressure in brachial artery(two-step method)
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Pre-Procedural Guidelines for Assessing Blood Pressure
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...