Related Experiment Video
Updated: Mar 21, 2026

Electrochemical Etching and Characterization of Sharp Field Emission Points for Electron Impact Ionization
Published on: July 12, 2016
Design of a 300-kV gas environmental transmission electron microscope equipped with a cold field emission gun
Shigeto Isakozawa1, Isao Nagaoki2, Akira Watabe2
1Hitachi High Technologies Corporation, 882 Ichige, Hitachinaka, Ibaraki 312-8504, Japan shigeto.isakozawa.rj@hitachi-hightech.com.
Abstract:
A new in situ environmental transmission electron microscope (ETEM) was developed based on a 300 kV TEM with a cold field emission gun (CFEG). Particular caution was taken in the ETEM design to assure uncompromised imaging and analytical performance of the TEM. Because of the improved pumping system between the gun and column, the vacuum of CFEG was largely improved and the probe current was sufficiently stabilized to operate without tip flashing for 2-3 h or longer. A high brightness of 2.5 × 10(9) A/cm(2) sr was measured at 300 kV, verifying the high quality of the CFEG electron beam. A specially designed gas injection-heating holder was used in the in situ TEM study at elevated temperatures with or without gas around the TEM specimen. Using this holder in a 10 Pa gas atmosphere and specimen temperatures up to 1000°C, high-resolution ETEM performance and analysis were achieved.
Related Concept Videos
Transmission Electron Microscopy
Overview of Electron Microscopy
Preparation of Samples for Electron Microscopy
Scanning Electron Microscopy
Fundamental Principles
Accelerated...