A New Cuffless Device for Measuring Blood Pressure: A Real-Life Validation Study
Tessa S Schoot1, Mariska Weenk, Tom H van de Belt
1Radboud University Medical Center, Department of Internal Medicine, Radboud University, Nijmegen, Netherlands.
Cuffless blood pressure (BP) monitors offer a convenient alternative to traditional cuffs. The Checkme Health Monitor showed promising correlation with reference monitors, particularly when calibrated in the supine position.
Area of Science:
- Biomedical Engineering
- Cardiovascular Technology
- Health Monitoring Devices
Background:
- Cuffless blood pressure (BP) monitoring devices utilize pulse transit time for easier, faster, and cheaper BP measurement.
- These devices present a viable alternative to traditional cuff-based BP monitors for self-measurement.
Purpose of the Study:
- To evaluate the real-world performance of the Checkme Health Monitor, a novel cuffless BP monitor.
- To investigate the impact of volunteer posture and device position relative to heart level on BP measurement accuracy.
Main Methods:
- The study involved 37 volunteers with varying BP ranges, adhering to the European Society of Hypertension International Protocol.
- Systolic BP was measured using the Checkme device and a reference monitor (RM) in randomized supine and sitting positions.
- The influence of device placement at and above heart level was also assessed.
Main Results:
- BP results from the Checkme device correlated well with the reference monitor.
- In the supine position, significant differences (>5 mmHg) were observed in 46% of participants, with 14% showing differences >15 mmHg.
- Calibration position (supine) showed better correlation.
Conclusions:
- The Checkme cuffless BP monitor demonstrated promising correlation with a reference device, especially when calibrated in the supine position.
- Preliminary findings support further research into cuffless BP monitoring in clinical and outpatient settings.
More Related Videos
09:38Femoral Arterial and Venous Catheterization for Blood Sampling, Drug Administration and Conscious Blood Pressure and Heart Rate Measurements
Published on: January 24, 2012
05:51Measuring the Carotid to Femoral Pulse Wave Velocity Cf-PWV to Evaluate Arterial Stiffness
Published on: May 3, 2018
Related Concept Videos
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Measurement of Blood Pressure
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Pre-Procedural Guidelines for Assessing Blood Pressure
Assessment of blood pressure in brachial artery(two-step method)