Related Experiment Video
Updated: Mar 21, 2026

Author Spotlight: Translational Applications of Stimulated SFEMG in Rodent Models
Published on: March 8, 2024
The effect of different low-frequency filters on concentric needle jitter in stimulated orbicularis oculi
Anjla Patel1,2, Magdalena Gosk1,3, Matthew Pitt1
1Department of Clinical Neurophysiology, Great Ormond Street Hospital For Children NHS Foundation Trust, Great Ormond Street, London, WC1N 3JH, UK.
Introduction:
Our objective was to compare the effect of different low-frequency filters on jitter parameters when stimulating the orbicularis oculi.
Methods:
Ten healthy volunteers were studied. Jitter was expressed as the mean consecutive difference (MCD). The low filter settings compared were 1, 2, and 3 kHz.
Results:
No significant difference in mean MCD or outliers was found with the different filter settings. No significant difference in mean MCD was seen when the number of potentials analyzed was reduced.
Conclusion:
Different low-frequency settings do not influence the mean MCD when using a peak detection system. Muscle Nerve 54: 317-319, 2016.