Development of an In Vitro Ocular Platform to Test Contact Lenses
Published on: April 6, 2016
Short-term comparison between extended depth-of-focus prototype contact lenses and a commercially-available
Daniel Tilia1, Anna Munro2, Jiyoon Chung2
1The Brien Holden Vision Institute, Sydney, NSW 2052, Australia; School of Optometry and Vision Science, The University of New South Wales, Sydney, NSW 2052, Australia.
Purpose:
To compare the visual performance of prototype contact lenses which extend depth-of-focus (EDOF) by deliberate manipulation of multiple higher-order spherical aberration terms and a commercially-available center-near lens (AIR OPTIX Aqua Multifocal, AOMF).
Methods:
This was a prospective, cross-over, randomized, single-masked (participant), short-term clinical trial where 52 participants (age 45-70 years) were stratified as low, medium or high presbyopes and wore EDOF and AOMF on different days. Objective measures comprised high and low contrast visual acuity (HCVA/LCVA, logMAR), and contrast sensitivity (log units) at 6m; HCVA at 70cm, 50cm and 40cm and stereopsis (seconds of arc) at 40cm. HCVA at 70cm, 50cm and 40cm were measured as "comfortable acuity" rather than conventional resolution acuity. Subjective measures comprised clarity-of-vision and ghosting at distance, intermediate and near, overall vision satisfaction and ocular comfort (1-10 numeric rating scale) and lens purchase (yes/no response). Statistical analysis included repeated measures ANOVA, paired t-tests and McNemar's test.
Results:
Significant differences between lens types were independent of strata (p≥0.119). EDOF was significantly better than AOMF for HCVA at 40cm (0.42±0.18 vs. 0.48±0.22, p=0.024), stereopsis (98±88 vs. 141±114, p<0.001), clarity-of-vision at intermediate (8.5±1.6 vs. 7.7±1.9, p=0.006) and near (7.3±2.5 vs. 6.2±2.5, p=0.005), lack-of-ghosting (p=0.012), overall vision satisfaction (7.5±1.7 vs. 6.4±2.2, p<0.001) and ocular comfort (9.0±1.0 vs. 8.3±1.7, p=0.002). Significantly more participants chose to only-purchase EDOF (33% vs. 6%, p=0.003).). There were no significant differences between lens types for any objective measure at 6m or clarity-of-vision at distance (p≥0.356).
Conclusions:
EDOF provides better intermediate and near vision performance in presbyopes than AOMF with no difference for distance vision during short-term wear.
More Related Videos
05:46Author Spotlight: Advancements in Refractive Surgical Correction for Presbyopia and Exploring Postoperative Visual Acuity
Published on: September 20, 2024
05:14Comparison of Agreement and Accuracy using Binocular Wavefront Optometer with Autorefractor and Phoropter
Published on: September 16, 2025
Related Concept Videos
Focusing of Light in the Eye
Depth Perception and Spatial Vision