Related Experiment Video
Updated: Mar 21, 2026

High-resolution Structural Magnetic Resonance Imaging of the Human Subcortex In Vivo and Postmortem
Published on: December 30, 2015
Toward 20 T magnetic resonance for human brain studies: opportunities for discovery and neuroscience rationale
Thomas F Budinger1, Mark D Bird2, Lucio Frydman2,3
1Lawrence Berkeley National Laboratory, University of California, Berkeley, CA, USA. tfbudinger@lbl.gov.
Abstract:
An initiative to design and build magnetic resonance imaging (MRI) and spectroscopy (MRS) instruments at 14 T and beyond to 20 T has been underway since 2012. This initiative has been supported by 22 interested participants from the USA and Europe, of which 15 are authors of this review. Advances in high temperature superconductor materials, advances in cryocooling engineering, prospects for non-persistent mode stable magnets, and experiences gained from large-bore, high-field magnet engineering for the nuclear fusion endeavors support the feasibility of a human brain MRI and MRS system with 1 ppm homogeneity over at least a 16-cm diameter volume and a bore size of 68 cm. Twelve neuroscience opportunities are presented as well as an analysis of the biophysical and physiological effects to be investigated before exposing human subjects to the high fields of 14 T and beyond.
Related Concept Videos
Magnetic Resonance Imaging
Brain Imaging
These technologies include computerized axial tomography (CAT or CT scans), positron-emission tomography (PET scans), magnetic resonance imaging (MRI), functional magnetic resonance imaging (fMRI), and Transcranial Magnetic...
Applications Of NMR In Biology