Comparison of cytotoxicity test models for evaluating resin-based composites
Human & Experimental Toxicology
|May 21, 2016
Summary
Cytotoxicity testing models for resin-based composites (RBCs) showed consistent results. Bulk-fill RBCs demonstrated biocompatibility at 4mm, indicating safe clinical application.
Area of Science:
- Biomaterials Science
- Dental Materials
- Toxicology
Background:
- Resin-based composites (RBCs) are widely used in dentistry.
- Evaluating the biocompatibility of RBCs is crucial for patient safety.
- Different cytotoxicity testing models exist, but their consistency needs evaluation.
Purpose of the Study:
- To compare different cytotoxicity test models for RBCs.
- To assess the biocompatibility of standard and bulk-fill RBCs.
- To evaluate the effectiveness of cure in relation to cytotoxicity.
Main Methods:
- Standard (spectrum TPH) and bulk-fill (SDR) RBCs were tested.
- Specimens of 2 and 4 mm thickness were prepared and polymerized.
- Micro-hardness and three cytotoxicity tests (direct, indirect, extract) using L-929 cells were performed.
Main Results:
- Lower hardness and higher cytotoxicity were observed for TPH at 4mm.
- SDR demonstrated biocompatibility at 4mm across all tested models.
- Strong correlations were found between hardness ratio, cell viability, and the different test models.
Conclusions:
- Direct, indirect, and extract cytotoxicity test models showed consistency for RBC evaluation.
- Bulk-fill RBCs are biocompatible when placed and cured at 4mm thickness.
- The study validates the reliability of these cytotoxicity testing methods for dental materials.
Related Concept Videos
In Vitro Drug Dissolution: Compendial Testing Models I
458
Compendial dissolution methods are standardized procedures defined by pharmacopeias to evaluate the rate at which a drug dissolves in a specific medium. These methods ensure batch-to-batch consistency, enable quality control, and support the prediction of drug bioavailability. They are critical for both immediate and modified-release drug products.The apparatuses used for dissolution testing differ in their design and mechanical function, but all aim to simulate the physiological environment of...
458
In Vitro Drug Dissolution: Compendial Testing Models II
495
Various dissolution methods are utilized to assess a drug’s dissolution rate, including the flow-through cell, paddle-over-disk, cylinder, and reciprocating disk methods.The flow-through cell apparatus (USP (United States Pharmacopeia) method 4) comprises a reservoir for the dissolution medium and a pump that propels the medium through the cell containing the test sample. This method is crucial for assessing modified-release dosage forms with minimally soluble active ingredients,...
495