[Evaluation of Measurement Accuracy and Inter-institutional Comparison for Dose Calibrators]
Norikazu Matsutomo1, Yoshiharu Kangai, Nobuhiro Yada
1Department of Radiological Technology, Kurashiki Central Hospital (Current address: Department of Medical Radiological Technology, Faculty of Health Sciences, Kyorin University).
Purpose:
The aim of this study was to validate the reliability of dose calibrators for measuring the radioactivity of several radioisotopes in multi-institution.
Methods:
We evaluated the measurement accuracy of dose calibrators using a commercially available source ((67) Ga, (99m) Tc, (123) I, (201) TL). Nine dose calibrators (five models) in seven institutions were performed in this study. Each source was measured at least 3 times a day over a period of 4 half-life. Linearity of concentration (%error value) and percent difference values (%diff measurement) between measured and estimated radioactivity were calculated to evaluate the measurement accuracy. In addition, difference among institutions (%diff institution) was evaluated by the error values between measured and reference institution values.
Results:
Good linearity of concentration was found between measured and estimated radioactivity in (99m)Tc and (123)I. However, %error value was increased in (67)Ga and (201)TL (maximum 19.3%). %diff measurements were 1.9 ± 0.3% for (67)Ga, -0.9 ± 0.3% for (99m)Tc, 2.2 ± 0.4% for (123)I, and -0.7 ± 0.3% for (201)TL, respectively. Although there were no clear differences in six institutions, %diff institution in one institution tended to be higher than that obtained in other institutions.
Conclusions:
Our results indicated that measurement accuracy of nine dose calibrators (five models) was relatively stable. However, difference of measured values tended to be higher in a part of institution and source. It is important to perform quality assurance and quality control for dose calibrator using traceable source.
More Related Videos
Related Concept Videos
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Glassware Calibration
Volumetric flasks: Volumetric flasks are designed to prepare aqueous solutions of precise volumes accurately with a calibration line on the neck. To calibrate a volumetric flask, it is important to fill it with distilled...
Accuracy and Precision
Uncertainty in Measurement: Reading Instruments
Uncertainty in Measurement: Accuracy and Precision
Dose Size and Dosing Frequency: Determination Methods