An accurate method for real-time chest compression detection from the impedance signal
Heemun Kwok1, Jason Coult2, Chenguang Liu3
1Center for Progress in Resuscitation, University of Washington, Seattle, WA, United States; Department of Medicine, University of Washington School of Medicine, Seattle, WA, United States.
Objective:
Real-time feedback improves CPR performance. Chest compression data may be obtained from an accelerometer/force sensor, but the impedance signal would serve as a less costly, universally available alternative. The objective is to assess the performance of a method which detects the presence/absence of chest compressions and derives CPR quality metrics from the impedance signal in real-time at 1s intervals without any latency period.
Methods:
Defibrillator recordings from cardiac arrest cases were divided into derivation (N=119) and validation (N=105) datasets. With the force signal as reference, the presence/absence of chest compressions in the impedance signal was manually annotated (reference standard). The method classified the impedance signal at 1s intervals as Chest Compressions Present, Chest Compressions Absent or Indeterminate. Accuracy, sensitivity and specificity for chest compression detection were calculated for each case. Differences between method and reference standard chest compression fractions and rates were calculated on a minute-to-minute basis.
Results:
In the validation set, median accuracy was 0.99 (IQR 0.98, 0.99) with 2% of 1s intervals classified as Indeterminate. Median sensitivity and specificity were 0.99 (IQR 0.98, 1.0) and 0.98 (IQR 0.95, 1.0), respectively. Median chest compression fraction error was 0.00 (IQR -0.01, 0.00), and median chest compression rate error was 1.8 (IQR 0.6, 3.3) compressions per minute.
Conclusion:
A real-time method detected chest compressions from the impedance signal with high sensitivity and specificity and accurately estimated chest compression fraction and rate. Future investigation should evaluate whether an impedance-based guidance system can provide an acceptable alternative to an accelerometer-based system.
Related Concept Videos
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Instrumentation Amplifier
To overcome this challenge, an ECG machine utilizes an instrumentation amplifier. This specialized amplifier is...
Bode Plots Construction