Influence of automatic frequent pace-timing adjustments on effective left ventricular pacing during cardiac
Niraj Varma1, Robert W Stadler2, Subham Ghosh2
1Cleveland Clinic Main Campus, Mail Code J2-2, 9500 Euclid Ave., Cleveland, OH 44195, USA.
Aims:
Cardiac resynchronization therapy (CRT) requires effective left ventricular (LV) pacing (i.e. sufficient energy and appropriate timing to capture). The AdaptivCRT™ (aCRT) algorithm serves to maintain ventricular fusion during LV or biventricular pacing. This function was tested by comparing the morphological consistency of ventricular depolarizations and percentage effective LV pacing in CRT patients randomized to aCRT vs. echo-optimization.
Methods And Results:
Continuous recordings (≥20 h) of unipolar LV electrograms from aCRT (n = 38) and echo-optimized patients (n = 22) were analysed. Morphological consistency was determined by the correlation coefficient between each beat and a template beat. Effective LV pacing of paced beats was assessed by algorithmic analysis of negative initial EGM deflection in each evoked response. The %CRT pacing delivered, %effective LV pacing (i.e. % of paced beats with effective LV pacing), and overall %effective CRT (i.e. product of %CRT pacing and %effective LV pacing) were compared between aCRT and echo-optimized patients. Demographics were similar between groups. The mean correlation coefficient between individual beats and template was greater for aCRT (0.96 ± 0.03 vs. 0.91 ± 0.13, P = 0.07). Although %CRT pacing was similar for aCRT and echo-optimized (median 97.4 vs. 98.6%, P = 0.14), %effective LV pacing was larger for aCRT [99.6%, (99.1%, 99.9%) vs. 94.3%, (24.3%, 99.8%), P=0.03]. For aCRT vs. echo-optimized groups, the proportions of patients with ≥90% effective LV pacing was 92 vs. 55% (P = 0.002), and with ≥90% effective CRT was 79 vs. 45%, respectively (P = 0.018).
Conclusion:
AdaptivCRT™ significantly increased effective LV pacing over echo-optimized CRT.
Related Concept Videos
Heart Failure VI: Adjunct Therapies
Heart Failure Drugs: Inotropic Agents
Imbalances in Cardiac Output
CHF can occur due to the failure of either side of the heart. Left-side failure leads to pulmonary congestion—the right side continues to send...
Cardiac Output I:Effect of Heart Rate on Cardiac Output
Cardiac output (CO) refers to the total amount of blood ejected by one of the ventricles in liters per minute (L/min). In a resting adult, CO ranges from 5 to 6 L/min, adjusting according to the body's metabolic requirements.
Effect of Heart Rate on Cardiac Output
Cardiac output adapts to metabolic demands during stress, physical activity, or illness. The autonomic nervous system regulates heart rate via the sinoatrial node. The parasympathetic nervous system decreases heart...
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Heart Failure II: Pathophysiology