In Vitro Comparison of Second- and Third-generation Dual-source CT for Coronary Stent Visualization at Different Tube
Tobias Gassenmaier1, Nils Petri2, Thomas Allmendinger3
1Department of Diagnostic and Interventional Radiology, University Hospital of Würzburg, Oberdürrbacher Str. 6, 97080 Würzburg, Germany.
Rationale And Objectives:
The study aimed to evaluate in vitro stent lumen visibility of coronary stents in a second- and third-generation dual-source computed tomography (CT) system at 100 and 120 kVp tube potential.
Materials And Methods:
Twenty-six coronary stents ranging from 2.25 to 4.0 mm in diameter were implanted in a coronary vessel phantom. Scans were performed at 100 and 120 kVp tube potential. Evaluation was performed using a medium-sharp kernel in both systems (B46f in the second-generation and Bv49 in the third-generation model) and a sharp (Bv59) convolution kernel optimized for vascular imaging in the third-generation CT.
Results:
The median visible stent lumen diameter in the second-generation system was higher at 120 kVp with a median of 62.0% compared to 56.3% at 100 kVp (P < 0.001). The median visible diameter in the third-generation system was significantly higher applying the Bv49 kernel with 66.7% at 120 kVp and 61.1% at 100 kVp (both P < 0.001). When applying the Bv59 kernel, visible stent lumen further increased to 69.3% at 120 kVp and 66.7% at 100 kVp. Additionally, stent lumen was assessed using full width at half maximum, resulting in a comparable increase in luminal diameter at corresponding tube potential.
Conclusions:
Third-generation dual-source CT provides superior stent lumen visibility at equivalent tube potential and at reduced tube potential of 100 kVp when compared to 120 kVp in a second-generation system, at least when manually assessed.
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System II:Types of Echocardiography
Types of Echocardiography
Transthoracic Echocardiography (TTE)
TTE is the most common type of echocardiogram which involves placing a transducer on the patient's chest, emitting sound waves to create heart images. TTE is invaluable for evaluating the heart's size, structure, and motion, making it particularly useful for...
Imaging Studies III: Gastrointestinal Motility Studies and Virtual Colonoscopy
Radionuclide Testing
Radionuclide testing is a sophisticated medical technique for assessing gastrointestinal motility. It focuses on gastric emptying and colonic transit time. Radioactive markers track the movement of food through the digestive system, providing insights into gastrointestinal disorders.
In gastric emptying studies, a meal's liquid and...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...