Effect of photon energy spectrum on dosimetric parameters of brachytherapy sources
Mahdi Ghorbani1, Mohammad Mehrpouyan2, David Davenport3
1Medical Physics Research Center, Mashhad University of Medical Sciences, Mashhad, Iran.
Aim:
The aim of this study is to quantify the influence of the photon energy spectrum of brachytherapy sources on task group No. 43 (TG-43) dosimetric parameters.
Background:
Different photon spectra are used for a specific radionuclide in Monte Carlo simulations of brachytherapy sources.
Materials And Methods:
MCNPX code was used to simulate 125I, 103Pd, 169Yb, and 192Ir brachytherapy sources. Air kerma strength per activity, dose rate constant, radial dose function, and two dimensional (2D) anisotropy functions were calculated and isodose curves were plotted for three different photon energy spectra. The references for photon energy spectra were: published papers, Lawrence Berkeley National Laboratory (LBNL), and National Nuclear Data Center (NNDC). The data calculated by these photon energy spectra were compared.
Results:
Dose rate constant values showed a maximum difference of 24.07% for 103Pd source with different photon energy spectra. Radial dose function values based on different spectra were relatively the same. 2D anisotropy function values showed minor differences in most of distances and angles. There was not any detectable difference between the isodose contours.
Conclusions:
Dosimetric parameters obtained with different photon spectra were relatively the same, however it is suggested that more accurate and updated photon energy spectra be used in Monte Carlo simulations. This would allow for calculation of reliable dosimetric data for source modeling and calculation in brachytherapy treatment planning systems.
More Related Videos
Related Concept Videos
Biological Effects of Radiation
Dual Nature of Electromagnetic (EM) Radiation
Wavelength is the distance between two consecutive peaks (the highest point) or troughs (the lowest point) in the wave. Frequency is the number of...
Interaction of EM Radiation with Matter: Spectroscopy
Absorption of Radiation
Atomic Absorption Spectroscopy: Radiation and Light Sources
Two common narrow-range 'line' sources used in AAS are hollow-cathode lamps (HCLs) and...
Mutations
Chromosomal Alterations Are Large-Scale Mutations
While point mutations are changes in a single nucleotide in...