On-demand calibration and evaluation for electromagnetically tracked laparoscope in augmented reality visualization
Xinyang Liu1, William Plishker2, George Zaki2
1Sheikh Zayed Institute for Pediatric Surgical Innovation, Children's National Health System, 111 Michigan Avenue NW, Washington, DC, 20010, USA.
Summary
This study introduces fCalib, a rapid calibration method for electromagnetically (EM) tracked laparoscopes, enabling on-demand operating room calibration. The method achieves accuracy comparable to existing techniques in under 23 seconds.
Area of Science:
- Medical Imaging
- Surgical Technology
- Augmented Reality
Background:
- Current laparoscopic augmented reality systems require lengthy camera calibration procedures.
- Traditional methods necessitate multiple images of a checkerboard pattern, hindering real-time operating room (OR) use.
- There is a need for a fast, on-demand calibration solution for laparoscopes in the OR.
Purpose of the Study:
- To develop a rapid calibration method for electromagnetically (EM) tracked laparoscopes.
- To enable on-demand laparoscope calibration directly within the operating room.
- To reduce the time and complexity associated with current camera calibration techniques.
Main Methods:
- A mechanical tracking mount was designed to position an EM sensor on a conventional laparoscope.
- A tool named fCalib was developed for calibrating intrinsic and extrinsic camera parameters using a single image of a special target pattern.
- A tube phantom was integrated for rapid evaluation of calibration results in the OR.
Main Results:
- The fCalib method demonstrated comparable spatial target registration error to the OpenCV method in laboratory settings.
- Re-projection accuracy was evaluated between EM and optical tracking-based fCalib in a clinical setting.
- The fCalib method's computational time averaged 14.0 seconds, with a range of 3.5 to 22.7 seconds.
Conclusions:
- A prototype for fast calibration and evaluation of EM-tracked laparoscopes was successfully developed and validated.
- The developed calibration method is accurate, fast, and suitable for on-demand use in the OR.
- The fCalib method offers a practical solution for improving augmented reality systems in laparoscopic surgery.
Keywords:
Augmented realityCamera calibrationElectromagnetic trackingLaparoscopic procedureLaparoscopic visualization