Evaluation of intraocular lens position during phacoemulsification using intraoperative spectral-domain
Lyubomyr M Lytvynchuk1, Carl G Glittenberg1, Christiane I Falkner-Radler1
1From the Department of Ophthalmology (Lytvynchuk, Glittenberg, Falkner-Radler, Neumaier-Ammerer, Smretschnig, Hagen, Ansari-Shahrezaei, Binder), Rudolf Foundation Hospital, the Karl Landsteiner Institute for Retinal Research and Imaging (Lytvynchuk, Glittenberg, Falkner-Radler, Neumaier-Ammerer, Smretschnig, Hagen, Ansari-Shahrezaei, Binder), and the Retina Center (Ansari-Shahrezaei, Binder), Vienna, and the Department of Ophthalmology (Ansari-Shahrezaei), Medical University of Graz, Graz, Austria; the Ophthalmology Department (Lytvynchuk), University Clinic Gießen and Marburg GmbH, Gießen, Germany; the Professor Sergienko Eye Clinic (Lytvynchuk), Vinnytsia, Ukraine.
Purpose:
To assess the position of intraocular lenses (IOLs) at the end of standard phacoemulsification with intraoperative spectral-domain optical coherence tomography (SD-OCT).
Settings:
Department of Ophthalmology, Rudolf Foundation Hospital, Vienna, Austria.
Design:
Prospective case series.
Methods:
Standard phacoemulsification with IOL implantation was performed. The Rescan 700 SD-OCT system was used for intraoperative imaging. The anterior segment of the eye was scanned using SD-OCT at the end of the surgery. The distance from the IOL optic center and the IOL optic edge to the posterior capsule was measured postoperatively using graphic software.
Results:
The study comprised 74 patients (101 eyes). The mean axial length was 23.97 mm (range 21.43 to 28.61 mm). The mean IOL power was 20.39 diopters (D) (range 6.5 to 27.5 D). Contact between the IOL and posterior capsule was absent in 88 cases (87.13%), and partial or full contact was present in 13 cases (12.87%). The mean distance between the IOL central optic and posterior capsule was 0.71 pixel (range 0.06 to 1.38 pixels) in 99 cases (98.02%). In 42 cases (57.53%), partial contact between the IOL edges and the posterior capsule was noticed. The mean distance between the IOL edge and posterior capsule was 0.21 pixel (range 0.04 to 0.92 pixel).
Conclusions:
Intraoperative SD-OCT facilitated the imaging of IOL position during standard phacoemulsification. Contact between the IOL central optic and posterior capsule at the end of the surgery occurred rarely. Improved IOL design should be considered.
Financial Disclosure:
Drs. Binder and Glittenberg are consultants to Carl Zeiss Meditech AG. None of the other authors has a financial or proprietary interest in any material or method mentioned.
More Related Videos
07:44In vivo Structural Assessments of Ocular Disease in Rodent Models using Optical Coherence Tomography
Published on: July 24, 2020
04:59Author Spotlight: Enhancing Visual Outcomes in Cataract Surgery: A Novel Technique to Prevent Posterior Capsular Opacification Through IOL Rotation
Published on: July 7, 2023
Related Concept Videos
Angle Closure Glaucoma: Treatment
Open Angle Glaucoma: Treatment
Drugs such as carbonic anhydrase inhibitors, α2- and...