Calibration of Contactless Pulse Oximetry
Wim Verkruysse1, Marek Bartula, Erik Bresch
1From the *Standardization Research, †Patient Care and Measurements, and ‡Data Science, Philips Research, Eindhoven, The Netherlands.
Background:
Contactless, camera-based photoplethysmography (PPG) interrogates shallower skin layers than conventional contact probes, either transmissive or reflective. This raises questions on the calibratability of camera-based pulse oximetry.
Methods:
We made video recordings of the foreheads of 41 healthy adults at 660 and 840 nm, and remote PPG signals were extracted. Subjects were in normoxic, hypoxic, and low temperature conditions. Ratio-of-ratios were compared to reference SpO2 from 4 contact probes.
Results:
A calibration curve based on artifact-free data was determined for a population of 26 individuals. For an SpO2 range of approximately 83% to 100% and discarding short-term errors, a root mean square error of 1.15% was found with an upper 99% one-sided confidence limit of 1.65%. Under normoxic conditions, a decrease in ambient temperature from 23 to 7°C resulted in a calibration error of 0.1% (±1.3%, 99% confidence interval) based on measurements for 3 subjects. PPG signal strengths varied strongly among individuals from about 0.9 × 10 to 4.6 × 10 for the infrared wavelength.
Conclusions:
For healthy adults, the results present strong evidence that camera-based contactless pulse oximetry is fundamentally feasible because long-term (eg, 10 minutes) error stemming from variation among individuals expressed as A*rms is significantly lower (<1.65%) than that required by the International Organization for Standardization standard (<4%) with the notion that short-term errors should be added. A first illustration of such errors has been provided with A**rms = 2.54% for 40 individuals, including 6 with dark skin. Low signal strength and subject motion present critical challenges that will have to be addressed to make camera-based pulse oximetry practically feasible.
More Related Videos
09:33Using Near-Infrared Spectroscopy Wearable Devices to Identify Central Versus Peripheral Limitations During Exercise
Published on: December 19, 2024
07:51Standardized Hemorrhagic Shock Induction Guided by Cerebral Oximetry and Extended Hemodynamic Monitoring in Pigs
Published on: May 21, 2019
Related Concept Videos
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Special considerations while measuring pulse
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Instrument Calibration
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...