Multiplexed Isothermal Amplification Based Diagnostic Platform to Detect Zika, Chikungunya, and Dengue 1
Published on: March 13, 2018
A Simple Modification to the Mosquito Homogenization Protocol Safely Inactivates West Nile Virus and Allows Virus
Kristen L Burkhalter1, Brad J Biggerstaff1, Kalanthe Horiuchi1
1Division of Vector-Borne Diseases, Centers for Disease Control and Prevention, 3156 Rampart Road, Fort Collins, CO 80521.
Abstract:
We evaluated the ability of the Rapid Analyte Measurement Platform (RAMP(®)) mosquito-grinding buffer to inactivate West Nile virus (WNV) by subjecting WNV-positive samples ground in RAMP buffer to incubation intervals ranging from 5 min to 60 min. At each time point an aliquot was removed and serially diluted in bovine albumin (BA)-1 cell culture media to stop the inactivation process by RAMP buffer. Each BA-1 sample was tested for viable virus using Vero 6-well cell culture plaque assay and observed for plaques. We observed very limited inactivation of WNV (1-2 log10 plaque-forming units/ml) by RAMP buffer. Concerned for RAMP operators who may be using this assay in low-level biocontainment facilities, we developed an alternate sample homogenization protocol using Triton X-100 detergent that ensures complete WNV inactivation without compromising the performance of the RAMP assay.