ACA-Pro: calibration protocol for quantitative diffuse reflectance spectroscopy. Validation on contact and noncontact
Veronica Sorgato1, Michel Berger1, Charlotte Emain1
1CEA-LETI, Minatec Campus, 17 Rue des Martyrs, F38054 Grenoble, Cedex 9, France.
Journal of Biomedical Optics
|June 10, 2016
Summary
We created an adaptive calibration algorithm and protocol (ACA-Pro) to accurately measure tissue optical properties using diffuse reflectance spectroscopy (DRSsr). This method ensures reliable quantification across various systems and experimental conditions.
Area of Science:
- Biomedical Optics
- Spectroscopy
- Medical Imaging
Background:
- Diffuse reflectance spectroscopy (DRSsr) systems require accurate calibration to quantify tissue optical properties.
- Instrumental variations and experimental conditions can affect the reliability of DRSsr measurements.
Purpose of the Study:
- To develop and validate an adaptive calibration algorithm and protocol (ACA-Pro) for spatially resolved DRSsr.
- To enable robust quantification of absorption and scattering properties in biological tissues.
Main Methods:
- Developed ACA-Pro using a Monte Carlo-based look-up-table approach.
- Utilized a calibration reference base with various intralipid phantoms.
- Incorporated a solid material measurement for accounting experimental variations.
- Validated ACA-Pro in contact and noncontact probe-based DRSsr systems.
Main Results:
- ACA-Pro effectively corrects instrumental response in DRSsr systems.
- A single calibration reference base is applicable across different experiments.
- Validated robustness with a probe-free setup using a CCD detector.
Conclusions:
- ACA-Pro provides a reliable method for quantifying tissue optical properties.
- The protocol enhances the accuracy and consistency of DRSsr measurements.
- Demonstrated the adaptability and robustness of ACA-Pro for future applications.
More Related Videos
Related Concept Videos
Instrument Calibration
1.1K
Instrument calibration is essential for ensuring that instruments produce accurate and consistent results. It is vital in manufacturing, healthcare, testing laboratories, and scientific research. Calibration processes are specific to each instrument and help enhance data accuracy. Each instrument has a unique calibration process tailored to its design and function to improve data accuracy.
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
Analytical Balance Calibration
An analytical balance measures mass and requires regular calibration to...
1.1K
Attenuated Total Reflectance (ATR) Infrared Spectroscopy: Overview
1.5K
Attenuated total reflectance (ATR) infrared spectroscopy is a powerful analytical technique used to study the composition of materials. It is widely employed in chemistry, materials science, forensic science, and other fields where sample characterization is required. ATR has several advantages over traditional transmission IR spectroscopy, including the requirement of little to no sample preparation and the ability to analyze a wide range of samples.
The ATR process begins by directing a beam...
The ATR process begins by directing a beam...
1.5K
Spectrophotometry: Introduction
11.0K
Spectrophotometry is the quantitative measurement of the absorption, reflection, diffraction, or transmission of electromagnetic radiation through a material as a function of the intensity and wavelength of the radiation. A spectrophotometer is a device used to measure the change in the radiation intensity caused by its interaction with the material.
The essential components of a spectrophotometer include a source of electromagnetic radiation, a slot for placing a material to be analyzed, and a...
The essential components of a spectrophotometer include a source of electromagnetic radiation, a slot for placing a material to be analyzed, and a...
11.0K