Related Experiment Video
Updated: Mar 19, 2026

The Muscle Cuff Regenerative Peripheral Nerve Interface for the Amplification of Intact Peripheral Nerve Signals
Published on: January 13, 2022
Unacceptable failure of the PI2® implant
J van Aaken1, N Holzer1, L Wehrli2
11 Service de chirurgie orthopédique et traumatologie de l'appareil moteur, Hôpital Universitaire de Genève, Genève, Switzerland.
Abstract:
The PI2® spacer is designed for treatment of trapeziometacarpal (TM) osteoarthritis. However, the shape of this implant has raised concerns about its stability. We retrospectively investigated 45 implants in 41 patients treated for trapeziometacarpal osteoarthritis in our hospital between 2004 and 2009 who underwent trapeziectomy and insertion of a PI2® spacer. Outcome parameters included revision rates and clinical outcomes correlated with implant position and scaphometacarpal distance, assessed using standard radiographs. A total of 12 implants (27%) were removed at a median time of 10 months (interquartile range (IQR), 7-22 months). These included five dislocations and one early infection. Additionally, a further six patients underwent revision due to persistent pain. Three of these had scapho-trapezoid osteoarthritis, two had developed subluxation of the implant, and one did not show any radiographic abnormalities. A review of patient records revealed that 33 implants remained in place at a median time of 29 months (IQR, 20-57). However, of those, only 21 implants (64%) in 17 patients were available for clinical evaluation at a median follow-up of 29 months (IQR, 19-62 months). No significant differences in clinical outcomes including functional results were observed between in-place ( n = 8) and subluxated ( n = 13) implants. Due to the high revision rate (12/45), consistent with other reports in the literature, we have abandoned the use of the PI2® spacer. We recommend the establishment of a registry for evaluation of future implants.
Type Of Study/Level Of Evidence:
Case-series study/level IV.
More Related Videos
06:39Novel Percutaneous Approach for Deployment of 3D Printed Coronary Stenosis Implants in Swine Models of Ischemic Heart Disease
Published on: February 18, 2020
11:49Autologous Endothelial Progenitor Cell-Seeding Technology and Biocompatibility Testing For Cardiovascular Devices in Large Animal Model
Published on: September 9, 2011