Objective:
The aim of the current study was to compare 180° and 360° data collection modes to measure end diastolic volume (EDV), end systolic volume (ESV) and ejection fraction (EF) values of the cardiac system by gated myocardial perfusion tomography.
Methods:
Thirty-three patients underwent gated myocardial perfusion tomography. Single photon emission computed tomography data of patients' heart were acquired by 180°, 45° left posterior oblique to 45° right anterior oblique, and 360° to obtain EDV, ESV, EF and cardiac volume changes (V1, V2, V3, V4, V5, V6, V7 and V8) throughout each cardiac cycle.
Results:
Results of the current study indicated that there were no significant differences between 180° and 360° angular sampling in terms of measuring EDV, ESV and EF in myocardial perfusion imaging. Cardiac volume change patterns during a cardiac cycle were also similar in 360° and 180° scans. We also observed that there was no difference in EDV, ESV and EF values between the group with stress induced by exercise and the group with stress imposed by dipyridamole.
Conclusion:
As there is no difference between 180°and 360° cardiac scanning in terms of EDV, ESV and EF, half-orbit scan is recommended to study these cardiac system parameters because it offers more comfort to patients and a shorter scanning time.
More Related Videos
11:09High-Resolution Cardiac Positron Emission Tomography/Computed Tomography for Small Animals
Published on: December 16, 2022
08:13In Vivo Quantitative Assessment of Myocardial Structure, Function, Perfusion and Viability Using Cardiac Micro-computed Tomography
Published on: February 16, 2016
Related Concept Videos
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Imaging Studies for Cardiovascular System V: CT