Low contrast detectability and spatial resolution with model-based Iterative reconstructions of MDCT images: a
Domitille Millon1, Alain Vlassenbroek2, Aline G Van Maanen3
1Department of Radiology and Medical Imaging, Cliniques Universitaires Saint Luc, Université Catholique de Louvain, Avenue Hippocrate 10, 1200, Brussels, Belgium. domitille.millon@uclouvain.be.
Objectives:
To compare image quality [low contrast (LC) detectability, noise, contrast-to-noise (CNR) and spatial resolution (SR)] of MDCT images reconstructed with an iterative reconstruction (IR) algorithm and a filtered back projection (FBP) algorithm.
Methods:
The experimental study was performed on a 256-slice MDCT. LC detectability, noise, CNR and SR were measured on a Catphan phantom scanned with decreasing doses (48.8 down to 0.7 mGy) and parameters typical of a chest CT examination. Images were reconstructed with FBP and a model-based IR algorithm. Additionally, human chest cadavers were scanned and reconstructed using the same technical parameters. Images were analyzed to illustrate the phantom results.
Results:
LC detectability and noise were statistically significantly different between the techniques, supporting model-based IR algorithm (p < 0.0001). At low doses, the noise in FBP images only enabled SR measurements of high contrast objects. The superior CNR of model-based IR algorithm enabled lower dose measurements, which showed that SR was dose and contrast dependent. Cadaver images reconstructed with model-based IR illustrated that visibility and delineation of anatomical structure edges could be deteriorated at low doses.
Conclusion:
Model-based IR improved LC detectability and enabled dose reduction. At low dose, SR became dose and contrast dependent.
Key Points:
• Model- based Iterative Reconstruction improves detectability of low contrast object. • With model- based Iterative Reconstruction, spatial resolution is dose and contrast dependent. • Model-based Iterative Reconstruction algorithms enable improved IQ combined with dose-reduction possibilities. • Improvement of SR and LC detectability on the same IMR data set would reduce reconstructions.
More Related Videos
12:15Tissue Preparation Techniques for Contrast-Enhanced Micro Computed Tomography Imaging of Large Mammalian Cardiac Models with Chronic Disease
Published on: February 8, 2022
10:26Author Spotlight: A 3D Digital Model for the Diagnosis and Treatment of Pulmonary Nodules
Published on: May 19, 2023
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...