[Comparison and optimization of total ionic strength adjustment buffer during detecting fluoride in trace serum
Objective:
To probe on the influence by four kinds of total ionic strength regulating buffer (TISAB) in establishing the standard curve and determining the level of fluoride in trace serum sample (rats serum and the calf serum) by fluoride ion selective electrode, and further explore the optimal volume ratio among TISAB and serum samples.
Methods:
Standard curve equations of four TISAB were established to clarify the influence of TISAB kinds. And the trace serum samples (rats and calf) were diluted to 1 times by TISAB, and then levels of fluoride in the samples that added by the standard sodium fluoride were detected, subsequently the influence of TISAB kinds on recovery rate were analyzed. Finally, the optimal ratio among serum sample and each TISAB in different serum was determined according to the recovery rate of the serum sample with different volume ratio.
Results:
The standard curve of TISAB III sample exhibited best fitting effect (R2 = 0.9926 ), and it also showed the overwhelming superiority with the scope of recovery rates ranged from 90% to 105% in the rat serum standard addition recovery experiments. As calf serum, TISAB IV (recommended by health industry standard) performed superior than others with the scope of recovery rate ranged from 90% to 110% in standard addition recovery experiment. When the volume ratio changed, every kinds of TISAB had an optimal proportion with good recovery rate, only TISAB III with the proportion as 1: 1, nevertheless, those change little affected on the recovery rate in calf serum.
Conclusion:
The method recommended by health industry standard is only suitable for detecting the trace amounts of fluoride in the serum sample as calf, but not the serum sample as rats.
Related Concept Videos
EDTA: Indirect and Alkalimetric Titration
Calculating pH Changes in a Buffer Solution
Effects of EDTA on End-Point Detection Methods
In the visual method, metal-ion indicators (metallochromic dyes), which have distinct colors in their free and complex forms, are added to the mixture to signal the titration's end point. They form stable complexes with metal ions, but these complexes are weaker than the corresponding metal–EDTA complexes. As a...
Complexometric EDTA Titration Curves
EDTA: Direct, Back-, and Displacement Titration
Direct titration involves buffering the metal ion solution to the desired pH and directly titrating with standard EDTA until the endpoint. The optimum pH ensures a large conditional formation constant of metal−EDTA and visibility of the free indicator color in the solution. In addition, auxiliary complexing reagents are used to prevent the precipitation of metal hydroxides...
Precipitation Titration: Endpoint Detection Methods
In the Volhard method, a standard excess of AgNO3 is first added to the...