Protein Complex Affinity Capture from Cryomilled Mammalian Cells
Published on: December 9, 2016
Contamination risk of stable isotope samples during milling
M Isaac-Renton1,2, L Schneider2, K Treydte2
1Department of Renewable Resources, University of Alberta, 751 General Services Building, Edmonton, Alberta, T6G 2H1, Canada.
Rationale:
Isotope analysis of wood is an important tool in dendrochronology and ecophysiology. Prior to mass spectrometry analysis, wood must be homogenized, and a convenient method involves a ball mill capable of milling samples directly in sample tubes. However, sample-tube plastic can contaminate wood during milling, which could lead to biological misinterpretations.
Methods:
We tested possible contamination of whole wood and cellulose samples during ball-mill homogenization for carbon and oxygen isotope measurements. We used a multi-factorial design with two/three steel milling balls, two sample amounts (10 mg, 40 mg), and two milling times (5 min, 10 min). We further analyzed abrasion by milling empty tubes, and measured the isotope ratios of pure contaminants.
Results:
A strong risk exists for carbon isotope bias through plastic contamination: the δ(13) C value of polypropylene deviated from the control by -6.77‰. Small fibers from PTFE filter bags used during cellulose extraction also present a risk as the δ(13) C value of this plastic deviated by -5.02‰. Low sample amounts (10 mg) showed highest contamination due to increased abrasion during milling (-1.34‰), which is further concentrated by cellulose extraction (-3.38‰). Oxygen isotope measurements were unaffected.
Conclusions:
A ball mill can be used to homogenize samples within test tubes prior to oxygen isotope analysis, but not prior to carbon or radiocarbon isotope analysis. There is still a need for a fast, simple and contamination-free sample preparation procedure. Copyright © 2016 John Wiley & Sons, Ltd.