Compliance charts to guide non-complex small artery stenting: validation by quantitative coronary angiography
Yohan Chacko1, J Kimberly Haladyn2, Debbie M Smith3
1Department of Medicine, The University of Queensland, Brisbane, Queensland, Australia; Department of Cardiology, Princess Alexandra Hospital, Brisbane, Queensland, Australia.
Objective:
To determine whether stent sizing derived from manufacturers' compliance charts provides a reasonable in vivo estimate of final minimum lumen diameter (MLD) when compared with quantitative coronary angiography (QCA).
Design:
Single-centre measurement comparison study.
Setting:
Tertiary referral university hospital.
Patients:
Fifty cases receiving a single stent for non-complex de novo stenosis were randomly selected from the percutaneous coronary intervention database of our high-volume centre. Restenosis, stent thrombosis, bifurcational disease, rotablation, left main or graft stenting, intravascular ultrasound or kissing balloon inflations were exclusion criteria.
Main Outcome Measures:
Equality and limits of agreement (LOA) between compliance chart and QCA measurements of final MLD, especially focusing on patients with small stents<3 mm. The paired t test and Bland-Altman plots were used to compare measurements.
Results:
There was no significant difference between compliance chart-derived and QCA final MLD (n=50; mean -0.034 mm, SD 0.35, 95% CI -0.132 to +0.064; p=0.49), with reasonable Bland-Altman LOA between the two methods of assessing final MLD in the overall group (LOA -0.72 to +0.66 mm), as well as in the group of particular interest with Derived final MLD <3 mm (n=30; mean 0.019 mm, SD 0.27, 95% CI -0.082 to +0.119; p=0.71; LOA -0.52 to +0.56 mm).
Conclusions:
Compliance charts provide an acceptable estimate of final MLD and are a reasonable guide to sizing during non-complex stenting, especially in small vessels <3 mm.