Related Experiment Video
Updated: Mar 19, 2026

Real-Time DC-dynamic Biasing Method for Switching Time Improvement in Severely Underdamped Fringing-field Electrostatic MEMS Actuators
Published on: August 15, 2014
The Development of a Dual-Warhead Impact System for Dynamic Linearity Measurement of a High-g
Yunbo Shi1,2, Zhicai Yang3,4, Zongmin Ma5,6
1Science and Technology on Electronic Test and Measurement Laboratory, North University of China, Taiyuan 030000, Shanxi, China. shiyunbo@nuc.edu.cn.
Abstract:
Despite its extreme significance, dynamic linearity measurement for high-g accelerometers has not been discussed experimentally in previous research. In this study, we developed a novel method using a dual-warhead Hopkinson bar to measure the dynamic linearity of a high-g acceleration sensor with a laser interference impact experiment. First, we theoretically determined that dynamic linearity is a performance indicator that can be used to assess the quality merits of high-g accelerometers and is the basis of the frequency response. We also found that the dynamic linearity of the dual-warhead Hopkinson bar without an accelerometer is 2.5% experimentally. Further, we verify that dynamic linearity of the accelerometer is 3.88% after calibrating the Hopkinson bar with the accelerometer. The results confirm the reliability and feasibility of measuring dynamic linearity for high-g accelerometers using this method.
Related Concept Videos
Mass Analyzers: Common Types
Mass Analyzers: Overview