Externally Delivered Focused Ultrasound for Renal Denervation
Petr Neuzil1, John Ormiston2, Todd J Brinton3
1Department of Cardiology, Homolce Hospital, Prague, Czech Republic.
Objectives:
The aim of this study was to assess clinical safety and efficacy outcomes of renal denervation executed by an externally delivered, completely noninvasive focused therapeutic ultrasound device.
Background:
Renal denervation has emerged as a potential treatment approach for resistant hypertension.
Methods:
Sixty-nine subjects received renal denervation with externally delivered focused ultrasound via the Kona Medical Surround Sound System. This approach was investigated across 3 consecutive studies to optimize targeting, tracking, and dosing. In the third study, treatments were performed in a completely noninvasive way using duplex ultrasound image guidance to target the therapy. Short- and long-term safety and efficacy were evaluated through use of clinical assessments, magnetic resonance imaging scans prior to and 3 and 24 weeks after renal denervation, and, in cases in which a targeting catheter was used to facilitate targeting, fluoroscopic angiography with contrast.
Results:
All patients tolerated renal denervation using externally delivered focused ultrasound. Office blood pressure (BP) decreased by 24.6 ± 27.6/9.0 ± 15.0 mm Hg (from baseline BP of 180.0 ± 18.5/97.7 ± 13.7 mm Hg) in 69 patients after 6 months and 23.8 ± 24.1/10.3 ± 13.1 mm Hg in 64 patients with complete 1-year follow-up. The response rate (BP decrease >10 mm Hg) was 75% after 6 months and 77% after 1 year. The most common adverse event was post-treatment back pain, which was reported in 32 of 69 patients and resolved within 72 h in most cases. No intervention-related adverse events involving motor or sensory deficits were reported. Renal function was not altered, and vascular safety was established by magnetic resonance imaging (all patients), fluoroscopic angiography (n = 48), and optical coherence tomography (n = 5).
Conclusions:
Using externally delivered focused ultrasound and noninvasive duplex ultrasound, image-guided targeting was associated with substantial BP reduction without any major safety signals. Further randomized, sham-controlled trials will be needed to validate this unique approach.
More Related Videos
08:35Improved Renal Denervation Mitigated Hypertension Induced by Angiotensin II Infusion
Published on: May 26, 2022
05:56Evaluation of the Feasibility, Safety, and Accuracy of an Intraoperative High-intensity Focused Ultrasound Device for Treating Liver Metastases
Published on: January 9, 2019
Related Concept Videos
Imaging Studies II: Ultrasonography
Urinary Tract Calculi VI: Surgical Management