[A stress-dependent pulsatile cardiovascular model as the basis for regulating artificial hearts]
Related Experiment Video
Updated: Aug 14, 2026

An Isolated Working Heart System for Large Animal Models
Published on: June 11, 2014
Abstract:
The development and simulation of closed-loop control of blood pumps requires a pulsatile model of the cardiovascular system that takes into account the natural mechanisms of adaptation during exercise. In this article, a model is described which takes account of the baroreceptor reflex, the change in heart rate, contractility of the heart, and the peripheral resistance during exercise, as well as venous tone, and the nonlinearity of vessel compliance. In addition, a model of the artificial heart is presented, in which the nonlinear limitation of the stroke volume is taken into account, and the concept of a pulse frequency modulated (PFM) control system for the circulation with an artificial heart is described.
Related Concept Videos
Pathophysiology of Cardiac Performance
Neural Regulation of Blood Pressure
Baroreceptor Reflex
Baroreceptors, located in the carotid sinuses and aortic arch, detect changes in blood pressure. When blood pressure rises, these stretch-sensitive receptors...
Autoregulation of Blood Flow
Chemical Signaling in Autoregulation
Chemical signaling operates at the precapillary sphincter level, inciting either contraction or relaxation.
Heart Failure II: Pathophysiology