Comparison of image quality and patient discomfort in prostate MRI: pelvic phased array coil vs. endorectal coil
Borna K Barth1, Alexander Cornelius2, Daniel Nanz1
1Institute of Diagnostic and Interventional Radiology, University Hospital Zurich and University of Zurich, Raemistrasse 100, 8091, Zurich, Switzerland.
Purpose:
To compare image quality (IQ) and patient discomfort during prostate MRI using a pelvic phased array (PPA) coil and an endorectal (ER) coil.
Materials And Methods:
Ninety-eight patients (median age, 65.7; range 42.1-78.1) underwent prostate MRI on a 3T scanner including T2w and DWI acquired with PPA and an ER coil within the same exam. Acquisition time was kept similar for both acquisitions. Two radiologists evaluated aspects of IQ on a 5-point Likert scale and classified image artifacts. All patients completed a questionnaire on discomfort/pain regarding the ER coil using a visual analogue scale from 1 to 10.
Results:
There was no significant difference in overall IQ for T2w images for both readers (reader 1, 3.27 ± 0.91 and 3.07 ± 0.84, p = 0.057; reader 2, 3.70 ± 0.75 and 3.77 ± 0.81, p = 0.555) for PPA and ER coils, respectively. Overall IQ for DWI acquired with PPA and ER coils was rated similar by reader 1 (3.03 ± 1.10 and 3.08 ± 0.80, respectively, (p = 0.67)), while reader 2 preferred ER coil images (3.27 ± 0.81 and 3.66 ± 0.85 (p < 0.05)). Susceptibility artifacts were more frequent in ER than in PPA coil images (109 vs. 75). Discomfort and pain experienced during insertion of the ER coil was low altogether (VAS score, 3.5 ± 2.1 for "discomfort" and 2.4 ± 2.4 for "pain").
Conclusion:
T2-weighted images may be acquired with comparable IQ using a PPA coil as compared to an ER coil, while DWI images showed better IQ using the ER coil for one of two readers. The insertion of the ER coil caused low to moderate discomfort and pain in patients.
Insights
Pelvic phased array (PPA) coils offer comparable image quality to endorectal (ER) coils for prostate MRI T2w images. DWI images showed improved quality with the ER coil for one reader, with minimal patient discomfort.
Area of Science:
- Radiology
- Medical Imaging
- Oncology
Background:
- Prostate magnetic resonance imaging (MRI) is crucial for diagnosis and staging.
- Traditional endorectal (ER) coils provide high-resolution images but can cause patient discomfort.
- Pelvic phased array (PPA) coils offer a non-invasive alternative, but their comparative performance needs evaluation.
Purpose of the Study:
- To compare image quality (IQ) and patient discomfort between pelvic phased array (PPA) coils and endorectal (ER) coils for prostate MRI.
- To assess the impact of coil choice on T2-weighted (T2w) and diffusion-weighted imaging (DWI) quality.
- To quantify patient-reported discomfort associated with ER coil use.
Main Methods:
- Ninety-eight patients underwent 3T prostate MRI using both PPA and ER coils for T2w and DWI sequences.
- Radiologists assessed IQ on a 5-point Likert scale and identified artifacts.
- Patients completed a visual analogue scale (VAS) questionnaire regarding ER coil discomfort and pain.
Main Results:
- No significant difference in overall T2w IQ between PPA and ER coils for either reader.
- Reader 2 rated DWI IQ higher with the ER coil (p < 0.05), while Reader 1 found similar IQ for both.
- Susceptibility artifacts were more frequent with ER coils (109 vs. 75).
- Mean VAS scores for ER coil discomfort and pain were low (3.5 ± 2.1 and 2.4 ± 2.4, respectively).
Conclusions:
- PPA coils provide comparable T2w image quality to ER coils in prostate MRI.
- ER coils may offer superior DWI quality for some readers, despite increased artifact potential.
- ER coil insertion results in low to moderate patient discomfort and pain, suggesting it remains a viable option.
Related Concept Videos
Magnetic Resonance Imaging
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies IV: Magnetic Resonance Imaging
Imaging Studies for Cardiovascular System IV: CMRI