IntegoTM infusion system: cost effectiveness analysis focusing on dosimetry, sterility and management
Federica Matteucci1, Laura Solinas2, Valentina Turri3
1Nuclear Medicine Unit, Istituto Scientifico Romagnolo per lo Studio e la Cura dei Tumori (IRST), Forlì-Cesena, Italy - federica.matteucci@irst.emr.it.
Background:
Healthcare providers across Europe are facing an ever-growing demand in clinical PET referrals. Currently, it is estimated that the administration of the PET tracer accounts for approximately 40% of the unitary PET procedure reimbursement (uPETr). Although the cost of PET/CT is highly dependent on the radiopharmaceutical cost itself, little is known about the economic impact of the utilized administration method and the repercussions on staff radiation exposure. Our objective was to evaluate the cost-effectiveness of automatic injection/fractionation system Intego™ (Bayer HealthCare, MEDRAD Europe, Netherlands) for istaff radiation exposure reduction and to validate its use with 18F-choline (FCH).
Methods:
In order to validate Intego™ use with FCH we analyzed sterility, radioactivity fractionation accuracy and radiation protection for staff. We analyzed Intego™ impact on examination costs and its impact on organization efficiency. A cost-effectiveness analysis (CEA) was estimated as the incremental cost to reduce staff radiationexposure.
Results:
According to our data, Intego™ ensures both sterility and accuracy of FCH doses' activity, reducing, at the same time, the exposure to radiation either whole body and at the extremities (94% and 75% respectively for the technicians and complete reduction for physicians). Intego™'s variable unit costs are higher than the SA (respectively 1.8% and 0.4% of PET reimbursement), while staff costs are significantly higher with SA (respectively 0.27% and 1.57% of unitary PET reimbursement [uPETr]). In our simulation, based on a 2,450 PET yearly output, the differential costs were slightly higher by using Intego™™ (+ 14%). The incremental cost-effectiveness ratio (ICER) was equal to 1.1, i.e. the healthcare provider pays an additional cost of 0.38% of uPETr to obtain a significant reduction of staff radiation exposure (-4.5 µS).
Conclusions:
Intego™, for its favorable results in terms of cost effectiveness, could be a useful tool in a nuclear medicine department, limiting the staff radiation exposure.
More Related Videos
Related Concept Videos
Determination of Multiple Dosing Parameters: Steady-State, Minimum and Maximum Concentrations
One-Compartment Model: IV Infusion
The one-compartment model for IV infusion uses mathematical equations to describe the rate of change in drug quantity in the body. At steady-state or infusion equilibrium, the drug input...
Drug Accumulation During Multiple Dosing: Intermittent IV Infusions
Pharmaceutical Alternatives: Stability-Related Therapeutic Nonequivalence
Parenteral Drug Delivery Systems: Injectables, Implants, and Infusion Devices
Cleaning, Sterilization, and Disinfection
Cleaning
The cleaning process usually involves using water with detergents or enzymatic cleaner and removing foreign material from objects and surfaces, including organic material such as body fluids or inorganic material like soil. Cleaning is performed before high-level disinfection and sterilization because foreign materials on the cover of the devices interfere with process...