Evaluation of an Automated System for Reading and Interpreting Disk Diffusion Antimicrobial Susceptibility Testing of
Evgeny A Idelevich1, Karsten Becker1, Janne Schmitz2
1Institute of Medical Microbiology, University Hospital Münster, Münster, Germany.
Abstract:
Results of disk diffusion antimicrobial susceptibility testing depend on individual visual reading of inhibition zone diameters. Therefore, automated reading using camera systems might represent a useful tool for standardization. In this study, the ADAGIO automated system (Bio-Rad) was evaluated for reading disk diffusion tests of fastidious bacteria. 144 clinical isolates (68 β-haemolytic streptococci, 28 Streptococcus pneumoniae, 18 viridans group streptococci, 13 Haemophilus influenzae, 7 Moraxella catarrhalis, and 10 Campylobacter jejuni) were tested on Mueller-Hinton agar supplemented with 5% defibrinated horse blood and 20 mg/L β-NAD (MH-F, Oxoid) according to EUCAST. Plates were read manually with a ruler and automatically using the ADAGIO system. Inhibition zone diameters, indicated by the automated system, were visually controlled and adjusted, if necessary. Among 1548 isolate-antibiotic combinations, comparison of automated vs. manual reading yielded categorical agreement (CA) without visual adjustment of the automatically determined zone diameters in 81.4%. In 20% (309 of 1548) of tests it was deemed necessary to adjust the automatically determined zone diameter after visual control. After adjustment, CA was 94.8%; very major errors (false susceptible interpretation), major errors (false resistant interpretation) and minor errors (false categorization involving intermediate result), calculated according to the ISO 20776-2 guideline, accounted to 13.7% (13 of 95 resistant results), 3.3% (47 of 1424 susceptible results) and 1.4% (21 of 1548 total results), respectively, compared to manual reading. The ADAGIO system allowed for automated reading of disk diffusion testing in fastidious bacteria and, after visual validation of the automated results, yielded good categorical agreement with manual reading.
More Related Videos
09:59Application of the Intelligent High-Throughput Antimicrobial Sensitivity Testing/Phage Screening System and Lar Index of Antimicrobial Resistance
Published on: July 21, 2023
11:25Preparation of a Blood Culture Pellet for Rapid Bacterial Identification and Antibiotic Susceptibility Testing
Published on: October 15, 2014