Single and double acquisition strategies for compensation of artifacts from eddy current and transient oscillation in balanced steady-state free precession
Related Experiment Video
Updated: Mar 17, 2026

Optimized Setup and Protocol for Magnetic Domain Imaging with In Situ Hysteresis Measurement
Published on: November 7, 2017
Hyun-Soo Lee1, Seung Hong Choi2, Sung-Hong Park1
1MRI Laboratory, Department of Bio and Brain Engineering, Korea Advanced Institute of Science and Technology, Daejeon, Korea.
Purpose:
To develop single and double acquisition methods to compensate for artifacts from eddy currents and transient oscillations in balanced steady-state free precession (bSSFP) with centric phase-encoding (PE) order for magnetization-prepared bSSFP imaging.
Theory And Methods:
A single and four different double acquisition methods were developed and evaluated with Bloch equation simulations, phantom/in vivo experiments, and quantitative analyses. For the single acquisition method, multiple PE groups, each of which was composed of N linearly changing PE lines, were ordered in a pseudocentric manner for optimal contrast and minimal signal fluctuations. Double acquisition methods used complex averaging of two images that had opposite artifact patterns from different acquisition orders or from different numbers of dummy scans.
Results:
Simulation results showed high sensitivity of eddy-current and transient-oscillation artifacts to off-resonance frequency and PE schemes. The artifacts were reduced with the PE-grouping with N values from 3 to 8, similar to or better than the conventional pairing scheme of N = 2. The proposed double acquisition methods removed the remaining artifacts significantly. The proposed methods conserved detailed structures in magnetization transfer imaging well, compared with the conventional methods.
Conclusion:
The proposed single and double acquisition methods can be useful for artifact-free magnetization-prepared bSSFP imaging with desired contrast and minimized dummy scans. Magn Reson Med 78:254-263, 2017. © 2016 International Society for Magnetic Resonance in Medicine.
More Related Videos
09:01Gain-compensation Methodology for a Sinusoidal Scan of a Galvanometer Mirror in Proportional-Integral-Differential Control Using Pre-emphasis Techniques
Published on: April 4, 2017
07:01Frequency Mixing Magnetic Detection Scanner for Imaging Magnetic Particles in Planar Samples
Published on: June 9, 2016
Related Concept Videos
Double Resonance Techniques: Overview
Spin decoupling is usually achieved by...
NMR Spectrometers: Radiofrequency Pulses and Pulse Sequences
Eddy Currents
Other major applications of eddy currents appear in metal detectors and the braking systems of trains and roller...
Magnetic Damping
If, however, the bob is a slotted metal plate, the magnet produces a much smaller effect. When a slotted metal plate enters the field, an emf is induced by the change in flux; however, it is less effective because the slots limit the...
NMR Spectrometers: Resolution and Error Correction
Atomic Nuclei: Larmor Precession Frequency