Interference between surgical magnetic drapes and pacemakers: an observational study comparing commercially available
Valerie Zaphiratos1, Hubert Chiasson2, Pierre Drolet3
1Department of Anesthesia, Maisonneuve-Rosemont Hospital, University of Montreal, 5415 Boul. de L'Assomption, Montreal, QC, H1T 2M4, Canada. valerie@zaphiratos.ca.
Background:
Magnetic fields may potentially interfere with the function of cardiovascular implantable electronic devices. Sterile magnetic drapes used to hold surgical instruments are often placed on the patient's thorax, and they are likely to interfere with the function of these devices.
Methods:
Thirty patients were recruited to compare a new prototype surgical magnetic drape (LT10G™ by Menodys) made with bottom-isolated ferrite magnets to the Covidien magnetic drape we used in a previous study. Twenty additional patients were recruited to compare the prototype drape with four commercially available surgical magnetic drapes.
Results:
Magnetic interference was found in 33 of the 50 total patients (70 %) when the Covidien drape was placed over the pacemaker. Of the 20 additional patients, 5 patients (25 %) displayed magnetic interference with a second type of surgical magnetic drape. A third magnetic drape caused interference in one patient (5 %), whereas a larger drape of the same model did not interfere in any patient. No patients demonstrated magnetic interference with the prototype drape.
Conclusion:
Bottom isolation of magnets in the prototype magnetic drape (LT10G™) used during surgery prevents magnetic interference in all patients when placed over the pacemaker. Three of the four commercially available magnetic drapes tested demonstrated magnetic interference. Flipping the prototype drape is not recommended as it may expose non-isolated magnets to the cardiovascular implantable electronic device.
More Related Videos
08:51Magnetic Resonance Imaging of Multiple Sclerosis at 7.0 Tesla
Published on: February 19, 2021
11:46Investigation of the Electrophysiological and Thermographic Safety Parameters of Surgical Energy Devices During Thyroid and Parathyroid Surgery in a Porcine Model
Published on: October 13, 2022
Related Concept Videos
Magnetic Resonance Imaging
Imaging Studies for Cardiovascular System IV: CMRI