Early experience with automated B-mode quality assurance tests
1Medical Physics Department, United Lincolnshire Hospitals NHS Trust, Lincoln LN2 5QY.
Ultrasound (Leeds, England)
|July 20, 2016
Summary
Automated image quality analysis effectively detects scanner performance changes in quality assurance programs. This method demonstrates efficacy in identifying faults, though further optimization is needed for routine fault detection.
Area of Science:
- Medical Imaging
- Quality Assurance
- Diagnostic Ultrasound
Background:
- Quality assurance (QA) programs are crucial for medical imaging but lack robust evidence of efficacy.
- Automated image analysis offers a potential solution for objective and consistent QA.
Purpose of the Study:
- To assess the efficacy of an automated image quality analysis method in a QA program.
- To determine if automated analysis can detect changes in scanner performance.
Main Methods:
- Analysis of test object images measuring lateral resolution, low contrast penetration, slice thickness, and grey-scale contrast/visibility.
- Investigation of known/suspected scanner faults and review of routine QA results.
Main Results:
- Automated analysis detected changes in performance variables corresponding to known/suspected scanner faults.
- Image shadowing affected resolution and grey-scale visibility; user-reported quality deterioration correlated with changes in slice thickness, lateral resolution, and grey-scale contrast.
- A probe failure was identified due to an unrecorded change in default settings.
Conclusions:
- The automated image quality analysis method provides evidence of efficacy for scanner performance testing.
- Further research is needed to optimize the method for prospective fault detection and reduce confounding factors.
Related Concept Videos
Detection of Gross Error: The Q Test
When one or more data points appear far from the rest of the data, there is a need to determine whether they are outliers and whether they should be eliminated from the data set to ensure an accurate representation of the measured value. In many cases, outliers arise from gross errors (or human errors) and do not accurately reflect the underlying phenomenon. In some cases, however, these apparent outliers reflect true phenomenological differences. In these cases, we can use statistical methods...
Automated Microbial Diagnostics
Automated diagnostic analyzers have transformed clinical microbiology by providing rapid and reliable methods for pathogen identification and antibiotic susceptibility testing. Among these systems, the Vitek 2 is widely used because it automates the traditionally labor-intensive processes of microbial identification (ID) and antibiotic susceptibility testing (AST), delivering standardized and timely results that are essential for effective patient care.Microbial Identification with ID CardsThe...