Related Experiment Video
Updated: Mar 17, 2026

Neuronavigation and Laparoscopy Guided Ventriculoperitoneal Shunt Insertion for the Treatment of Hydrocephalus
Published on: October 14, 2022
Hydrogel-coated ventricular catheters for high-risk patients receiving ventricular peritoneum shunt
Hao Xu1, Yimin Huang, Wei Jiao
1Department of Neurosurgery, Tongji Hospital, Tongji Medical College, Huazhong University of Science and Technology, Wuhan Department of Neurosurgery, Anhui Provincial Hospital, Hefei, Anhui, China.
Abstract:
Shunt infection is a morbid complication of cerebrospinal fluid (CSF) shunting. The catheters with a hydrophilic surface may impede bacterial adherence and thereby reduce catheter-related CSF infection.A retrospective study compared the occurrence of CSF infection related to use of either standard silastic catheters or hydrogel-coated ventricular catheters (Bioglide, Medtronic). The enrolment was available to neurosurgery patients undergoing shunt surgery from October 2012 to 2015 in two centers. The follow-up period was more than months.A total of 78 patients were included in the study. In 33 patients 35-cm hydrogel-coated ventricular peritoneum shunts (VPS) were used, and in remaining 45 patients 35-cm standard silastic VPS catheters were used. Infection occurred in 14 (17.9%) patients, including definite VPS-related CSF infection in 6 patients (7.7%) and probable infection in remaining 8 patients (10.3%). There was a significant difference found in patients with total infection between the two groups [RR (95% CI); 0.200 (0.050-0.803), P = 0.014]. Analysis of Kaplan-Meier curve estimates indicated significant statistical difference between the two catheter types in duration (log rank = 4.204, P < 0.05). Significant statistical differences were also found in the subgroups including previous CSF infection within 1 month (log rank = 4.391, P = 0.04), conversion of external ventricular drains to shunt (Log Rank = 4.520, P = 0.03), and hospital stay >1 month (log rank = 5.252, P = 0.02). There was no difference found between the two groups of the patients with other infections within 1 month. The follow-up period was of 36 months.The hydrogel-coated catheter is a safe and related to lower infection rates for high-risk patients who underwent shunt surgery.
Related Concept Videos
Peritoneal Dialysis II: Peritoneal Dialysis Systems and Complications
Hemodialysis II: Procedure and Complications