Quantitative MRI assessment of a novel direction modulated brachytherapy tandem applicator for cervical cancer at
Abraam S Soliman1, Alyaa Elzibak2, Harry Easton3
1Department of Medical Physics, Sunnybrook Health Sciences Centre, Toronto, Canada; Physical Sciences, Sunnybrook Research Institute, Toronto, Canada.
Background And Purpose:
The purpose of this work is to quantitatively investigate the artifacts and image distortions induced in the MR images by a recently proposed direction modulated brachytherapy (DMBT) tandem applicator prototype. This new MRI-compatible applicator allows better sparing of organs-at-risk (OAR) for cervical cancer patients, while providing conformal dose distributions to target volumes.
Materials And Methods:
Specific phantom and tools were designed and manufactured for this study. The phantom was filled with a tissue-like solution and MR images were acquired with clinical protocols as per GEC-ESTRO recommendations. Images were obtained at 6 different orientations that mimic possible clinical settings and full-width-at-half-maximum (FWHM) was recorded at multiple locations/angles. The accuracy of detecting the centerline of the tandem was assessed using a novel radial-fiducials mount.
Results:
FWHM from all line profiles at all angles and all orientations was 6.14±0.7mm (compared to 6mm of the actual DMBT tandem diameter). The in-plane spatial-shift observed at para-axial and para-sagittal views was less than 0.5mm.
Conclusions:
This work demonstrated that the novel DMBT tandem applicator prototype has minimal artifact in T2-weighted images employed in clinical practice, suggesting the applicator might be a good candidate for MRI-guided adaptive brachytherapy.