Related Experiment Video
Updated: Mar 17, 2026

Novel Triple-Loop Technique for Suturing TFCC Injuries without Transosseous Tunnel
Published on: May 23, 2025
Tourniquet Effectiveness When Placed Over the Joint Service Lightweight Integrated Suit Technology
Background:
Chemical, biological, radiological, and nuclear threats (CBRNs) are uncommon; however, Special Operations Forces (SOF) are likely at the highest risk for tactical exposure. In the event of exposure, SOF will rely on the Joint Service Lightweight Integrated Suit Technology (JSLIST) for survival. Doctrine dictates that a tourniquet should be applied over the JSLIST after a severe limb injury with hemorrhage. There is no evidence in the literature that the Combat Application Tourniquet (C-A-T), which is currently the most widely available tourniquet on the battlefield, can effectively occlude arterial blood flow when applied over the JSLIST. We hypothesized that C-A-T application over the JSLIST would be ineffective at occluding arterial blood flow in the lower extremity.
Materials And Methods:
Following institutional review board approval, 20 healthy volunteers were recruited to participate. All volunteers wore the G3 Combat Pant and they donned the JSLIST. First, an operating room pneumatic tourniquet (gold standard) was applied in the proximal thigh and inflated to 300mmHg. Distal arterial interrogation was performed by examination of distal pulses and noninvasive arterial plethysmography wave-form analysis. After a 1-hour recovery period, the C-A-T was applied and tightened. A double routing technique was used, with three 180° turns of the windlass. The same distal interrogation followed. Half of the volunteers had the pneumatic tourniquet applied first, and the other half had the C-A-T applied first.
Results:
All volunteers had palpable pulses at baseline despite a wide range in volunteer body mass index. Distal pulses were absent in all volunteers following inflation of the pneumatic tourniquet as well as tightening of the C-A-T. The observed difference between the mean amplitude of plethysmographic waveforms was not different.
Conclusion:
The C-A-T effectively occludes arterial flow in the lower extremity, even when applied over the JSLIST. This finding supports existing military doctrine for tourniquet application over the JSLIST in the nonpermissive CBRN environment to control extremity exsanguination.
Related Concept Videos
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...
Venous Thrombosis III: Interprofessional Care
Veins of Lower Limbs
Formed by the union of the medial and lateral plantar veins, the posterior tibial vein, rising through the calf muscle, assimilates the fibular vein. The anterior tibial vein, a superior extension of the foot's dorsalis pedis vein, merges with the posterior tibial vein at the...
Personal Protective Equipment
Arteries of Lower Limbs
PPE Use in Healthcare Settings I: Donning