Quantitative Magnetic Resonance Imaging of Skeletal Muscle Disease
Published on: December 18, 2016
Test-retest analysis of multiple 31 P magnetization exchange pathways using asymmetric adiabatic inversion
Bertrand Pouymayou1,2, Tania Buehler1, Roland Kreis1
1Department of Clinical Research and Department of Radiology, University of Bern, Switzerland.
Purpose:
A 31 P-MR inversion transfer (IT) method with a short adiabatic inversion pulse is proposed and its test-retest reliability was evaluated for two spectral fitting strategies.
Methods:
Assessment in a test-retest design (3 Tesla, vastus muscles, 12 healthy volunteers, 14 inversion times, 22 ms asymmetric adiabatic inversion pulse, adiabatic excitation); spectral fitting in Fitting Tool for Interrelated Arrays of Datasets (FitAID) and Java Magnetic Resonance User Interface (jMRUI); least squares solution of the Bloch-McConnell-Solomon matrix formalism including all 14 measured time-points with equal weighting.
Results:
The cohort averages of k[PCr→γ-ATP] (phosphocreatine, PCr; adenosine triphosphate, ATP) are 0.246 ± 0.050s-1 versus 0.254 ± 0.050s-1 , and k[Pi→γ-ATP] 0.086 ± 0.033s-1 versus 0.066 ± 0.034s-1 (average ± standard deviation, jMRUI versus FitAID). Coefficients of variation of the differences between test and retest are lowest (9.5%) for k[PCr→γ-ATP] fitted in FitAID, larger (15.2%) for the fit in jMRUI, and considerably larger for k[Pi→γ-ATP] fitted in FitAID (43.4%) or jMRUI (47.9%). The beginning of the IT effect can be observed with magnetizations above 92% for noninverted lines while inversion of the ATP resonances is better than -72%.
Conclusion:
The performance of the asymmetric adiabatic pulse allows an accurate observation of IT effects even in the early phase; the least squares fit of the Bloch-McConnell-Solomon matrix formalism is robust; and the type of spectral fitting can influence the results significantly. Magn Reson Med 78:33-39, 2017. © 2016 International Society for Magnetic Resonance in Medicine.
More Related Videos
13:56The Use of Magnetic Resonance Spectroscopy as a Tool for the Measurement of Bi-hemispheric Transcranial Electric Stimulation Effects on Primary Motor Cortex Metabolism
Published on: November 19, 2014
08:59Use of a Multi-compartment Dynamic Single Enzyme Phantom for Studies of Hyperpolarized Magnetic Resonance Agents
Published on: April 15, 2016
Related Concept Videos
Double Resonance Techniques: Overview
Spin decoupling is usually achieved by...
Insensitive Nuclei Enhanced by Polarization Transfer (INEPT)