Related Experiment Video
Updated: Mar 17, 2026

Author Spotlight: Advancements in Impedance Monitoring for Cochlear Implant Surgery
Published on: August 4, 2023
Use of a mid-scala and a lateral wall electrode in children: insertion depth and hearing preservation
Abdelhamid Benghalem1, Dzemal Gazibegovic2, Fatima Saadi3
1a ENT Department , CLINIQUE RACHIDI , Casablanca , Morocco.
Conclusions:
Atraumatic insertion of the HiFocusTM Mid-Scala (HFMS) electrode via the round window was successfully achieved in seven children. Residual hearing 6 months post-operatively was preserved to within 10 dB HL of the pre-operative audiogram at 500 Hz for six children, indicating minimal initial insertion trauma to the cochlea.
Objectives:
The objectives were to document the clinical experience and evaluate differences between HFMS and HiFocusTM 1j (HF1j) by means of insertion depth and hearing preservation results.
Method:
Nineteen children were prospectively recruited and consecutively implanted with the HF1j electrode (n = 12) or the HFMS electrode (n = 7) via the round window. Average median angular insertion depths and the amount of residual hearing preserved at 6 months post-operatively were compared between the two electrode groups.
Results:
The median angular insertion depth for the HF1j was 439° and for the HFMS 435°. Preservation of residual hearing at 500 Hz was assessed in seven HFMS subjects and 11 HF1j subjects. Based on the Skarzynski formula, three out of seven subjects (42%) in the HFMS group had their residual hearing completely preserved at 500 Hz. In the control group, no subjects had complete hearing preservation and five subjects had a complete loss of residual hearing.

