Accuracy of radiographic measurements for implant planning using cone-beam and helical computer tomography
Carl K K Leung1, Edmond H N Pow2, Thomas K L Li3
1Comprehensive Dental Care, Faculty of Dentistry, The University of Hong Kong, Hong Kong, China.
Aim:
The aim of the present study was to compare the accuracy of radiographic measurements for dental implants planning using cone-beam computed tomography (CBCT) and helical computed tomography (HCT).
Methods:
Six pig ribs were wrapped by putty impression material, with radiographic markers placed. Two CBCT and an HCT were taken following the standard protocols. Twenty-five locations were selected, with vertical and horizontal dimensions measured using the default software, as well as on the processed HCT films by a digital caliper. The actual dimensions of the ribs measured by the digital caliper served as the control. Differences between radiographic dimensions and the actual dimensions were tested by two-way analysis of variance.
Results:
No differences were found between measurements made by CBCT and HCT images using the default software (P > 0.05). However, both measurements were statistically-significantly lower than the control (P < 0.001), and the mean difference was 0.3 mm. Measurements made on HCT films were statistically-significantly greater than the control (P < 0.001), and the mean difference was 0.5 mm.
Conclusion:
The accuracy of CBCT and HCT are similar, and both are reliable tools for implant planning. It is preferable to perform the planning using default software, rather than making direct measurements on films.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies III: Computed Tomography
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...