Accuracy of torque-limiting devices: A comparative evaluation
Haydar Albayrak1, Hasan Onder Gumus2, Funda Tursun3
1Assistant Professor, Department of Prosthodontics, Faculty of Dentistry, Erciyes University, Kayseri, Turkey.
Statement Of Problem:
To prevent the loosening of implant screws, clinicians should be aware of the output torque values needed to achieve the desired preload. Accurate torque-control devices are crucial in this regard; however, little information is currently available comparing the accuracy of mechanical with that of electronic torque-control devices.
Purpose:
The purpose of this in vitro study was to identify and compare the accuracy of different types of torque-control devices.
Material And Methods:
Devices from 5 different dental implant manufacturers were evaluated, including 2 spring-type (Straumann, Implance) mechanical devices (MTLD), 2 friction-type (Biohorizons, Dyna) MTLDs, and 1 (Megagen) electronic torque-control device (ETLD). For each manufacturer, 5 devices were tested 5 times with a digital torque tester, and the average for each device was calculated and recorded. The percentage of absolute deviations from the target torque values (PERDEV) were calculated and compared by using 1-way ANOVA. A 1-sample t test was used to evaluate the ability of each device to achieve its target torque value within a 95% confidence interval for the true population mean of measured values (α=.05 for all statistical analyses).
Results:
One-way ANOVAs revealed statistically significant differences among torque-control devices (P<.001). ETLD showed higher PERDEVs (28.33 ±9.53) than MTLDs (P<.05), whereas PERDEVS of friction-type (7.56 ±3.64) and spring-type (10.85 ±4.11) MTLDs did not differ significantly. In addition, devices produced by Megagen had a significantly higher (P<.05) PERDEV (28.33 ±9.53) other devices, whereas no differences were found in devices manufactured by Biohorizons (7.31 ±5.34), Dyna (7.82 ±1.08), Implance (8.43 ±4.77), and Straumann (13.26 ±0.79). However, 1-sample t tests showed none of the torque-control devices evaluated in this study were capable of achieving their target torque values (P<.05).
Conclusions:
Within the limitations of this in vitro study, MTLDs were shown to be significantly more accurate than ETLDs. However, none of the torque-control devices evaluated were able to meet their target torque values successfully.
Related Concept Videos
Torque
Torque can be considered as the rotational counterpart to force. Since forces change the translational...
Net Torque Calculations
Screw: Problem Solving
To evaluate the...
Circular Shafts - Elastoplastic Materials
As torque on the...
Transmission Shafts: Problem Solving
Next, use bending moment diagrams for the shaft to...
Stress Concentrations in Circular Shafts