A new automatic blood pressure kit auscultates for accurate reading with a smartphone: A diagnostic accuracy study
Hongjun Wu1, Bingjian Wang, Xinpu Zhu
1Department of Cardiology, Yixing No. 2 People's Hospital Department of Cardiology, Huai'an First People's Hospital, Nanjing Medical University, Jiangsu Province Department of Neurology, Branch of Shanghai General Hospital Department of Cardiology, Shanghai General Hospital, Shanghai Jiaotong University School of Medicine, Shanghai, People's Republic of China.
Abstract:
The widely used oscillometric automated blood pressure (BP) monitor was continuously questioned on its accuracy. A novel BP kit named Accutension which adopted Korotkoff auscultation method was then devised. Accutension worked with a miniature microphone, a pressure sensor, and a smartphone. The BP values were automatically displayed on the smartphone screen through the installed App. Data recorded in the phone could be played back and reconfirmed after measurement. They could also be uploaded and saved to the iCloud. The accuracy and consistency of this novel electronic auscultatory sphygmomanometer was preliminarily verified here. Thirty-two subjects were included and 82 qualified readings were obtained. The mean differences ± SD for systolic and diastolic BP readings between Accutension and mercury sphygmomanometer were 0.87 ± 2.86 and -0.94 ± 2.93 mm Hg. Agreements between Accutension and mercury sphygmomanometer were highly significant for systolic (ICC = 0.993, 95% confidence interval (CI): 0.989-0.995) and diastolic (ICC = 0.987, 95% CI: 0.979-0.991). In conclusion, Accutension worked accurately based on our pilot study data. The difference was acceptable. ICC and Bland-Altman plot charts showed good agreements with manual measurements. Systolic readings of Accutension were slightly higher than those of manual measurement, while diastolic readings were slightly lower. One possible reason was that Accutension captured the first and the last korotkoff sound more sensitively than human ear during manual measurement and avoided sound missing, so that it might be more accurate than traditional mercury sphygmomanometer. By documenting and analyzing of variant tendency of BP values, Accutension helps management of hypertension and therefore contributes to the mobile heath service.
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(two-step method)
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Measurement of Blood Pressure
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure:
Sites for measuring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement