Performances of a protector against scattered radiation during intraoperative use of a C-arm fluoroscope
Ki Hyuk Sung1, Young-Jun Jung2, Soon-Sun Kwon3
1Department of Orthopaedic Surgery, Seoul National University Bundang Hospital, 300 Gumi-Dong, Bundang-Gu, Seongnam, Gyeonggi 13620, Korea.
Abstract:
The scattered radiation protector for mobile x-ray systems, Creative Valuable Protector-2, has been recently developed. However, there have been no studies investigating the effects of this device. We aim to investigate the effects of the scattered radiation protector on the equivalent doses from scattered radiation delivered to radiosensitive organs while simulating spine surgery using a C-arm fluoroscope. Chest and rando phantoms were used to simulate a patient and a surgeon in this study. The equivalent dose from scattered radiation to radiosensitive organs was measured in four different situations according to the use of the scattered radiation protector and the C-arm configuration. To compare the quality of the images with and without the scattered radiation protector, an acryl step phantom with five steps was used, and the contrast resolution of each step was calculated. The equivalent dose from the scattered radiation to the surgeon's eye, thyroid, and gonad decreased significantly by using the scattered radiation protector for both the Posteroanterior (PA) (p < 0.001) and Anteroposterior (AP) (p < 0.001) C-arm configurations. The installation of the scattered radiation protector also reduced the direct radiation dose to the chest phantom. A scattered map showed that scattered radiation doses decreased by approximately 50% for the PA configuration and 75% for the AP configuration by using the scattered radiation protector. Before and after installation of the scattered radiation protector, the contrast resolution of each adjacent step area was 0.025-0.404 and 0.216-0.421. The scattered radiation protector was effective in reducing not only the equivalent dose from scattered radiation to the surgeon's radiosensitive organs, but also the direct radiation dose to the patient. This was all achieved without decreasing the quality of the C-arm fluoroscopic images.
Related Concept Videos
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
X-ray Imaging
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Radiological Investigation I: X-ray and CT
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System V: CT