A robust conversion method of radioactivities between plastic and NaI scintillation well counters for long-term
Takeshi Hara1, Tadahiko Ishihara2, Takahiko Asano2
1Department of Intelligent Image Information, Graduate School of Medicine, Gifu University, 1-1 Yanagido, Gifu, 501-1194, Japan. takeshi.hara@mac.com.
Purpose:
The purpose of this study is to specify a simple procedure for a robust data conversion of radioactivity value between plastic scintillator (PL) and NaI scintillator (NaI) devices.
Materials And Methods:
The radioactivity estimate of 100 blood samples was measured by the two devices. The two radioactivities were plotted on the same graph. The least-squares method was applied to obtain the conversion function. The differences between the actual radioradioy (N) from the NaI device and the estimated radioactivity for NaI (N') from the PL device activity (P) were statistically analyzed.
Results:
N' was determined from P as N' = 4.45 P + 6.28 with high correlation (r = 0.997). The Bland-Altman analysis between N' and N showed no fixed bias and no proportional bias.
Conclusions:
A hundred blood samples using a fixed type of sample tubes and a fixed radionuclide may be required to set up the robust conversion function.
More Related Videos
Related Concept Videos
Radioactive Decay and Radiometric Dating
Radioactivity and Nuclear Equations
A nuclide of an element has a specific number of protons and...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Nuclear Stability
To hold positively charged protons together...
![Automated Radiochemical Synthesis of [18F]3F4AP: A Novel PET Tracer for Imaging Demyelinating Diseases](/_next/image?url=https%3A%2F%2Fcloudfront.jove.com%2FCDNSource%2Fteasers%2F55537.jpg&w=3840&q=50)