Tumor Hypoxia Assessment: In Vivo 3D Oxygen Imaging Through Electron Paramagnetic Resonance
Published on: February 14, 2025
Skeletal Muscle Oxygenation Measured by EPR Oximetry Using a Highly Sensitive Polymer-Encapsulated Paramagnetic
1Department of Radiology, Geisel School of Medicine at Dartmouth, Hanover, NH, 03755, USA. Huagang.Hou@Dartmouth.Edu.
Abstract:
We have incorporated LiNc-BuO, an oxygen-sensing paramagnetic material, in polydimethylsiloxane (PDMS), which is an oxygen-permeable, biocompatible, and stable polymer. We fabricated implantable and retrievable oxygen-sensing chips (40 % LiNc-BuO in PDMS) using a 20-G Teflon tubing to mold the chips into variable shapes and sizes for in vivo studies in rats. In vitro EPR measurements were used to test the chip's oxygen response. Oxygen induced linear and reproducible line broadening with increasing partial pressure (pO2). The oxygen response was similar to that of bare (unencapsulated) crystals and did not change significantly on sterilization by autoclaving. The chips were implanted in rat femoris muscle and EPR oximetry was performed repeatedly (weekly) for 12 weeks post-implantation. The measurements showed good reliability and reproducibility over the period of testing. These results demonstrated that the new formulation of OxyChip with 40 % LiNc-BuO will enable the applicability of EPR oximetry for long-term measurement of oxygen concentration in tissues and has the potential for clinical applications.
Related Concept Videos
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.