Hyperpolarized Xenon for NMR and MRI Applications
Published on: September 6, 2012
Hyperpolarized (129) Xe MRI using isobutene as a new quenching gas
Shota Hodono1, Hirohiko Imai1, Yukiko Yamauchi1
1Department of Medical Physics and Engineering, Division of Health Sciences, Graduate School of Medicine, Osaka University, Suita, Osaka, Japan.
Abstract:
The use of a quenching gas, isobutene, with a low vapor pressure was investigated to enhance the utility of hyperpolarized (129) Xe (HP Xe) MRI. Xenon mixed with isobutene was hyperpolarized using a home-built apparatus for continuously producing HP Xe. The isobutene was then readily liquefied and separated almost totally by continuous condensation at about 173 K, because the vapor pressure of isobutene (0.247 kPa) is much lower than that of Xe (157 kPa). Finally, the neat Xe gas was continuously delivered to mice by spontaneous inhalation. The HP Xe MRI was enhanced twofold in polarization level and threefold in signal intensity when isobutene was adopted as the quenching gas instead of N2 . The usefulness of the HP Xe MRI was verified by application to pulmonary functional imaging of spontaneously breathing mice, where the parameters of fractional ventilation (ra ) and gas exchange (fD ) were evaluated, aiming at future extension to preclinical studies. This is the first application of isobutene as a quenching gas for HP Xe MRI.