Related Experiment Video
Updated: Mar 16, 2026

Cone Beam Intraoperative Computed Tomography-based Image Guidance for Minimally Invasive Transforaminal Interbody Fusion
Published on: August 6, 2019
Intraoperative Computed Tomography Imaging for Navigated Laparoscopic Renal Surgery: First Clinical Experience
Tobias Simpfendörfer1, Claudia Gasch1, Gencay Hatiboglu1
11 Department of Urology, University of Heidelberg , Heidelberg, Germany .
Introduction:
Laparoscopic partial nephrectomy (LPN) remains challenging in endophytic and complex kidney tumors as the clear understanding of tumor location and spreading depends on a precise analysis of available imaging. The purpose of this study was to investigate navigated kidney surgery using intraoperative cone-beam computed tomography (CBCT) images in conjunction with a previously proposed method for augmented reality (AR) guidance for safe LPN.
Materials And Methods:
The concept proposed is based on using an intraoperative CBCT scan for (1) marker-based AR guidance for fast and reliable tumor access and (2) enhancement of real-time fluoroscopy images for accurate tumor resection. Workflow and accuracy of the system were assessed using a porcine kidney model. Ten patients with complex or endophytic tumor localization and R.E.N.A.L. Nephrometry Score of at least nine scheduled for LPN were included in this study. Patients received an intraoperative CBCT after marker placement. Defining the resection line was assisted by AR. In addition, fluoroscopy imaging for depth perception was used for assistance during dissection. Feasibility and performance were assessed by histopathological results, peri- and postoperative data.
Results:
Surgery was performed successfully and negative margins were found in all cases. Segmental branches of the renal artery shifted as much as 10 mm in the vertical and 11 mm in the sagittal axis intraoperatively compared to preoperative imaging. Fluoroscopy to intraoperative computed tomography image fusion enabled enhanced depth perception during dissection in all cases. Radiation dose area product was 4.8 mGym2.
Conclusions:
The application of the navigation system is feasible and allows for safe and direct access to complex or endophytic renal masses. Radiation limits the application to selected indications.
Related Concept Videos
Imaging Studies III: Computed Tomography
Imaging Studies VII: Vascular Imaging
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies V: Intravenous Urography and Retrograde Pyelography
Imaging Studies IV: Magnetic Resonance Imaging
Kidney Transplant II: Surgical Procedure