Non-tapered PTFE capillary as robust and stable nanoelectrospray emitter for electrospray ionization mass
Di-Qiong Jin1, Ying Zhu1, Qun Fang1,2
1Institute of Microanalytical Systems, Department of Chemistry, Zhejiang University, Hangzhou, 310058, China.
Rationale:
The successful implementation of electrospray ionization mass spectrometry (ESI-MS) largely relies on the physical and chemical properties of electrospray emitters. Current ESI emitters based on pulled glass or fused-silica capillaries suffer from complicated fabrication procedures and short lifetime because of clogging and breakage problems. ESI emitters with high robustness, easy operation, and low cost are in great demand.
Methods:
We proposed to directly use non-pulled and non-tapered polytetrafluoroethylene (PTFE) capillaries as nanoelectrospray emitters. We evaluated and studied the performance of PTFE emitters under different flow rates and ESI voltages. Their applications in direct infusion and liquid chromatography (LC)-MS analysis were achieved.
Results:
High stable electrospray (relative standard deviation of total ion current intensity <5%) can be obtained at flow rates from 100 nL/min to 1500 nL/min and at ESI voltages from 1.8 kV to 2.9 kV using non-tapered PTFE emitters. Different types of samples including small-molecule drugs, peptides, and proteins can be detected with high sensitivity. LC-MS analysis of two model peptides was successfully demonstrated.
Conclusions:
We have demonstrated that a non-tapered PTFE capillary can be a promising ESI emitter for both direct infusion and LC-MS analysis. We envision it could find broad applications in high-throughput MS analysis fields, such as drug screening and proteomics, where robust and reproducible operations are essentially required. Copyright © 2016 John Wiley & Sons, Ltd.
Related Concept Videos
Electrospray Ionization (ESI) Mass Spectrometry
ESI utilizes electrical energy to transfer ions from the liquid phase of the sample into the...
Capillary Electrophoresis: Instrumentation
Capillary Electrophoresis: Applications
Capillary zone electrophoresis (CZE) separates ionic components based on their electrophoretic mobility. It has been used to separate proteins, amino acids,...
Electrophoresis: Overview
There...